Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
There is a potential that one or more image series (i.e., all images within an image set) may be missing from an exam without a warning displayed in the viewer.
Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog #: 801188, UDI 00886799000588. Used in cardio-pulmonary bypass.
Terumo Cardiovascular Systems
An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.
Conformis iTotal Hip Replacement System: HBS-033-00xx-010101, (HIP-PATIENT SPECIFIC STEM LEFT)
Conformis
Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.
Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values
Conformis iTotal Hip Replacement System: HBS-033-00xx-020101, (HIP-PATIENT SPECIFIC STEM RIGHT)
Conformis
Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.
There is a potential for an unexpected increase in stimulation during InterStim programming with the A10 Clinician Application (on Medtronic's smart programmer).
Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
Affected products may have been manufactured with the catheter distal end hole too small
Affected products may have been manufactured with the catheter distal end hole too small
The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .
The heated tube sets were consistently leaking and detaching from the cassette.
AngioDynamics Solero Microwave Tissue Ablation (MTA) System: Solero Generator PG (1) Catalog Number:12740000 UPN: H78712740000 (2) Solero Generator US PG Catalog Number: 12740000US UPN: H78712740000US0 (3) Solero Generator Service Loaner PG Catalog No: 12740000L UPN: H78712740000L0 (4) Solero Generator Service Loaner US PG Catalog Number: 12740000USL UPN: H78712740000USL0
Angiodynamics Inc. (Navilyst Medical Inc.)
Updated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure
ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV28F
OriGen Biomedical
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
2008T BlueStar Machine Software v.2.72-Hemodialysis machine Model Number: 191126 - Product Usage: The 2008T hemodialysis machine is indicated for acute and chronic dialysis therapy in a healthcare facility.
Fresenius Medical Care Renal Therapies Group
Power Logic Board Fail Message on the 2008T BlueStar Machine with software version 2.72.
The OMNIBotics iBlock Cutting Guide does not lock onto the iBlock Saw Guide as intended and may loosen during use.
Incorrect raw material used by the supplier in the manufacturing of the screw component, which could potentially lead to corrosion.
The heated tube sets were consistently leaking and detaching from the cassette.
ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV19F
OriGen Biomedical
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.