One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

Jul 3, 2019 Infusion Pumps View Details β†’

Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.

May 24, 2019 Diagnostic Equipment Nationwide View Details β†’

AXIOM Artis dBA, Model Number 7555357

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details β†’

AXIOM Artis dFA, Model Number 7555373

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details β†’

Artis zee floor, Model Number 10094135

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details β†’

An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.

May 16, 2018 Diagnostic Equipment Nationwide View Details β†’

AXIOM Artis dTA, Model Number 7008605

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details β†’

An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.

May 16, 2018 Diagnostic Equipment Nationwide View Details β†’

Artis zee biplane, Model Number 10094141

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details β†’

An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.

May 16, 2018 Diagnostic Equipment Nationwide View Details β†’

Artis zeego, Model Number 10280959

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details β†’