The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (Âŋ 0.2ÂŋC for Genius 2 and Âŋ 0.3ÂŋC for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2ÂŋC for Genius 2 or +0.3ÂŋC for Genius 3.

Oct 28, 2019 Diagnostic Equipment Nationwide View Details →

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (Âŋ 0.2ÂŋC for Genius 2 and Âŋ 0.3ÂŋC for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2ÂŋC for Genius 2 or +0.3ÂŋC for Genius 3.

Oct 28, 2019 Diagnostic Equipment Nationwide View Details →