The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Medtronic Open Pivot Aortic Valve Graft, REF 502AG25, SIZE 25 mm
MEDTRONIC ATS MEDICAL
Medtronic received a complaint from China that the label on the box for an Open Pivot Heart Valve was labeled as Aortic, but the product was Mitral. The valve itself was correctly labeled as Mitral.
Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
Medtronic MiniMed
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the user does not exit the bolus calculator as designed or when a system alarm interrupts a bolus calculation. If the user delivers the bolus, this may lead to hypoglycemia or hyperglycemia.
Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope
Karl Storz Endoscopy
During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.
The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly.
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com
QUIDEL CARDIOVASCULAR
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.
MiniMed Paradigm 512 Insulin pump
Medtronic MiniMed
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
LemonPrep Tubes (MD0019-T) and Single use cups (MD0019-SUP)
Carroll-Baccari
Samples of 114gm tubes of Lemon Prep, collected during a Food and Drug Administration inspection on October 15, 2019 were tested and found to be contaminated with Burkholderia cepacia. Mavidon is recalling all products manufactured at their facility.
MiniMed Paradigm Insulin pump, model number MMT-712
Medtronic MiniMed
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
FlipCutter III Drill
T.A.G. MEDICAL PRODUCTS CORPORATION
There is potential for the blister tray to be cracked causing the sterile barrier to be compromised and potential for particulate within the sterile barrier.
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
Due to physical limitations of the treatment machine, the maximum field opening for Siemens Virtual Wedge fields recommended by the LINAC vendor will be more restricted in the wedged direction than the maximum field opening without a wedge. Delivering a larger field opening with a virtual wedge will not result in a wedged field shape over the entire field. Instead, a high dose plateau region will be the result at the toe side of the field beyond the recommended field size.
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
Medtronic MiniMed Paradigm Insulin pump, model number MMT-712E
Medtronic MiniMed
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
MiniMed Paradigm Model 511 Insulin Pump
Medtronic MiniMed
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.