The firm has identified that cross-reactivity testing may not meet the claims indicated in the Barbiturates Reagent Chemistry Information Sheet (CIS). The presence of certain compounds in urine may result in false positive or false negative results.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
VITROS Immunodiagnostic Products TT4 Reagent Pack, Product Code 8744468
Ortho Clinical Diagnostics
The specified reagent packs exhibit an increase in the frequency of unsuccessful calibrations.
The Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose.
Prismaflex System, Prismaflex Control Unit
Baxter Healthcare
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
Prismaflex System, Prismaflex Control Unit
Baxter Healthcare
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wired or wireless acquisition modules.
WELCH ALLYN, INC/MORTARA
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..
ELI PC. The ELI PC Service automatically converts resting ECG recordings collected with WAM PC patient cables and WebUpload into DICOM 12-lead ECG Waveform files.
WELCH ALLYN, INC/MORTARA
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Prismaflex System, Prismaflex Control Unit
Baxter Healthcare
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged.
Potentially counterfeit products in/and counterfeit packaging were imported to the U.S.
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Cases of PortexΒΏ Lumbar Puncture Tray Child, SKU #4826PL-20 had been mislabeled with invalid information. The product was correct, but the inner trays were labeled as 4824PL-20 instead of 4826PL-20.
ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499
ICU Medical
Firm identified several software issues through internal testing of its Cogent Hemodynamic Monitoring System. These issues may result in inaccurate monitoring data, unit shutdown, and/or loss of display of a secondary parameter.
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
ABL800 FLEX model number 393-800 and 393-801, UDI:(01)05700693938004
Radiometer Medical ApS
Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from cyanosis and low blood pressure to cardiac arrhythmias and severe impairment of cardiac function.
CapsoCAM Plus, UDI: 00867770000209
Capso Vision
Capsule Endoscopy System: a defect present in the capsule window may be stressed through the production process and in the final package, which could cause the housing of the capsule to crack and leak when ingested. Fluid from the patient could leak in to the capsule, which may damage components inside the capsule and the patient data may not be retrieved; the patient may need to repeat the exam.
Prismaflex System, Prismaflex Control Unit
Baxter Healthcare
Baxter is communicating safety information from the Prismaflex Operator's Manual regarding the proper use of connectors with the Prismaflex system.