Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results

Aug 7, 2020 Diagnostic Equipment Nationwide View Details β†’

Sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.

Aug 28, 2020 Diagnostic Equipment Nationwide View Details β†’

lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.

Aug 4, 2020 Surgical Instruments Nationwide View Details β†’

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Aug 25, 2020 Other Medical Devices Nationwide View Details β†’

The tubes may have been delivered to some users without labels. Users may not be able to identify the tubes without these labels. If samples stored or transported in tubes are tested on certain systems or come into contact with bleach they can produce cyanide gas. The solution in these tubes contains guanidine thiocynate and may put laboratory personnel at risk of exposure to cyanide gas when it comes into contact with bleach or reagents containing bleach.

Jun 3, 2020 Infusion Pumps Nationwide View Details β†’

The firm received a report of an atypical dataset being generated. When using the cross-localization feature and images from a modality that generates asymmetrical images, the cross-localization reference line may be inaccurately placed on any of the corresponding images that are open.

Sep 10, 2019 Diagnostic Equipment Nationwide View Details β†’

Incorrect expiration date included in test kit labeling and/or in Substrate component label.

Jul 16, 2020 Other Medical Devices Nationwide View Details β†’

COVID-19 Interpretive Software issues that may cause: 1) the risk of invalid or inconclusive tests/false negative and potentially increasing the retesting burden on customers. 2) false positive results due to improper vortexing.

Aug 6, 2020 Diagnostic Equipment Nationwide View Details β†’

Incorrect expiration date included in test kit labeling and/or in Substrate component label.

Jul 16, 2020 Other Medical Devices Nationwide View Details β†’

Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results

Aug 7, 2020 Surgical Instruments Nationwide View Details β†’

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Aug 25, 2020 Other Medical Devices Nationwide View Details β†’

Mislabeled: WallFlex Biliary RX Uncovered Stent System 10x60mm devices labeled as WallFlexTM Biliary RX Fully Covered Stent System RMV 10x60mm device

Aug 7, 2020 Implants & Prosthetics Nationwide View Details β†’

PROFEMUR Titanium and Cobalt Chrome modular necks, Item Numbers and descriptions: Item Number Description Description 2 PHAC1202 PROFEMUR NECK NEUTRAL SHORT COBALT CHROME PHAC1204 PROFEMUR NECK NEUTRAL LONG COBALT CHROME PHAC1212 PROFEMUR NECK A/R VAR/VAL 2 SHORT COBALT CHROME PHAC1214 PROFEMUR NECK A/R VAR/VAL 2 LONG COBALT CHROME PHAC1222 PROFEMUR NECK A/R VAR/VAL 1 SHORT COBALT CHROME PHAC1224 PROFEMUR NECK A/R VAR/VAL 1 LONG COBALT CHROME PHAC1232 PROFEMUR NECK A/R 8DG SHORT COBALT CHROME PHAC1234 PROFEMUR NECK A/R 8DG LONG COBALT CHROME PHAC1242 PROFEMUR NECK A/R 15DG SHORT COBALT CHROME PHAC1244 PROFEMUR NECK A/R 15DG LONG COBALT CHROME PHAC1252 PROFEMUR NECK VAR/VAL 8DG SHORT COBALT CHROME PHA01202 PROFEMUR NECK NEUTRALSHORT SHORT PHA01204 PROFEMUR NECK NEUTRAL LONG PHA01206 PROFEMUR NECK NEUTRAL X LONG PHA01212 PROFEMUR NECK A/R VAR/VAL 2 SHORT PHA01214 PROFEMUR NECK A/R VAR/VAL 2 LONG PHA01222 PROFEMUR NECK A/R VAR/VAL 1 SHORT PHA01224 PROFEMUR NECK A/R VAR/VAL 1 LONG PHA01232 PROFEMUR NECK 8DG A/R SHORT PHA01234 PROFEMUR NECK 8DG A/R LONG PHA01236 PROFEMUR NECK 8DG A/R X LONG PHA01242 PROFEMUR NECK 15DG A/R SHORT PHA01244 PROFEMUR NECK 15DG A/R LONG PHA01252 PROFEMUR NECK 8DG VAR/VAL SHORT PHA01254 PROFEMUR NECK 8DG VAR/VAL LONG PHA01256 PROFEMUR NECK 8DG VAR/VAL X LONG PHA01262 PROFEMUR NECK 15DG VAR/VAL SHORT PHA01264 PROFEMUR NECK 15DG VAR/VAL LONG Intended Use: MPO Total Hip Systems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

MicroPort Orthopedics

Class I - Dangerous

MicroPort Orthopedics Inc. is voluntarily recalling any existing inventory of PROFEMURΒΏ Titanium and Cobalt Chrome modular necks with previous package insert versions to replace the PROFEMURΒΏ Hip System Package Insert with the most recent revision (150803-8).

Jul 31, 2020 Implants & Prosthetics Nationwide View Details β†’

Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.

Aug 4, 2020 Infusion Pumps Nationwide View Details β†’