LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Inquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.
Irvine Biomedical Inc, a St. Jude Medical
Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay
NM/CT 850 Model Number: H3907AD
GE Healthcare
Rotor bearing screws were found loose on detector.
Reagents were shipped to customers after the Use by data indicated on its labeling.
Central Line Dressing Change Tray
Medline Industries
Potential of incomplete seals impacting the sterility of the kit.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
Discovery NM 630 Model # H3101RH
GE Healthcare
Rotor bearing screws were found loose on detector.
Speed Stitch Needle Cassette
ArthroCare
Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
EVOLVE STEM 8.5MM Model # 496S085
Wright Medical Technology
Lack of sterility assurance
Inquiry Steerable Diagnostic Catheter, REF: IBI-87002 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.
Irvine Biomedical Inc, a St. Jude Medical
Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay
Discovery NM670 Pro Model # H3100XB
GE Healthcare
Rotor bearing screws were found loose on detector.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
NM/CT 870 DR Model # H3100AS
GE Healthcare
Rotor bearing screws were found loose on detector.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.
Beckman Coulter
Due to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results.
NM/CT 870 CZT Model # H3906CW
GE Healthcare
Rotor bearing screws were found loose on detector.
Pointts sistema de ellmlnact!;o de verrugas OTC - Brazil Catalog Number: 1001-0280 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of skin tags.
OraSure Technologies
Unit boxes not properly sealed
Personelle OTC Skin Tag Remover, Canada Catalog Number: 1001-0314 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of skin tags.
OraSure Technologies
Unit boxes not properly sealed
Software: Syngo.via RT Image Suite software version VA30_SP02 with license T_RT_IMAGE_SUITE_PAT_MARK Intended Use: used as an advanced application on a SOM10 scanner SMN: 10496180
Siemens Medical Solutions USA
Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not recognized during the subsequent treatment planning process or remains undetected during the position verification at the treatment system (e.g. linear accelerator), this can potentially result in an incorrect dose or dose applied to the wrong location