BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001
Roche Diagnostics Operations
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
The expiration date on labeling is incorrect.
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
Contour VL Variable Length Ureteral Stent Set, UPN M006180155090, M006180156090, M006180156180, M006180156190
Boston Scientific
Trace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel.
Roche cobas 8000 modular analyzer series (i.e., cobas 8000 ISE module; cobas c 502, 701, and 702 modules; and cobas e 602 and 801 modules). Chemistry Analyzer (photometric discrete), for clinical use Part Number: 08763662001
Roche Diagnostics Operations
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
PORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
Smiths Medical ASD
A specific model and lot number of Bivona Cuffless FlexTend Tracheostomy Tube may have been incorrectly packaged with a V-neck flange instead of a straight flange.
Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570
Boston Scientific
Trace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel.
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
IceSeed" 1.5 Cryoablation Needle
Boston Scientific
Complaint trend regarding needle shaft gas leaks.
ScopeValet Guardian Disposable Endoscope Valves Set Distributed By Ruhof Corporation, 393 Sagamore Avenue, Mineola, NY 11501 Product Code: 345SVVO1
GA HEALTH COMPANY LIMITED
Suction button is reported sticky and/or broken during or after the procedure, may cause prolonged procedure
IcePearl" 2.1 CX 90ΒΏ Cryoablation Needle
Boston Scientific
Complaint trend regarding needle shaft gas leaks.
Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.
IceRod" 1.5 CX 90ΒΏ Cryoablation NeedleΒΏ
Boston Scientific
Complaint trend regarding needle shaft gas leaks.
Five IceSphere" Prostate Cryoablation Kit
Boston Scientific
Complaint trend regarding needle shaft gas leaks.