ðŸĨ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

The plastic film used to manufacture these Angiography Drape and Cesarean Birth Drape products is causing the drapes to stick, which may prevent the products from being unfolded and applied properly. Sold in single sterile and bulk non-sterile configurations.

May 21, 2021 Surgical Instruments Nationwide View Details →

Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.

May 13, 2021 Diagnostic Equipment Nationwide View Details →

Failure to comply with the performance standard for fluoroscopy systems. Per 21 CFR 1020.32(k)(6), the displayed AKR and cumulative air kerma shall not deviate from the actual values by more than Âŋ 35 percent. The dosage displayed on the generator console and table display (monitor) is higher than the actual delivered dose.

Mar 11, 2021 Diagnostic Equipment Nationwide View Details →

The plastic film used to manufacture these Angiography Drape and Cesarean Birth Drape products is causing the drapes to stick, which may prevent the products from being unfolded and applied properly. Sold in single sterile and bulk non-sterile configurations.

May 21, 2021 Surgical Instruments Nationwide View Details →
Class I - Dangerous

Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.

Jun 3, 2021 Implants & Prosthetics View Details →
Class I - Dangerous

Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.

Jun 3, 2021 Implants & Prosthetics View Details →

ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result

May 18, 2021 Diagnostic Equipment Nationwide View Details →

The plastic film used to manufacture these Angiography Drape and Cesarean Birth Drape products is causing the drapes to stick, which may prevent the products from being unfolded and applied properly. Sold in single sterile and bulk non-sterile configurations.

May 21, 2021 Surgical Instruments Nationwide View Details →

The plastic film used to manufacture these Angiography Drape and Cesarean Birth Drape products is causing the drapes to stick, which may prevent the products from being unfolded and applied properly. Sold in single sterile and bulk non-sterile configurations.

May 21, 2021 Surgical Instruments Nationwide View Details →

Software error VE10B -In case of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION), the C-arm may leave its intended travel path by up to 22 cm (9") result in canceling clinical treatment or to continue treatment on an alternative system

May 27, 2021 Diagnostic Equipment Nationwide View Details →

Revolution CT

GE Healthcare

Class I - Dangerous

GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.

Mar 19, 2021 Other Medical Devices Nationwide View Details →