ðŸĨ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.

Jun 30, 2021 Other Medical Devices Nationwide View Details →

The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control certification statement [21 CFR 1010.2]. Philips has determined that some systems at customer locations did not have these labels applied.

Sep 22, 2021 Other Medical Devices Nationwide View Details →

BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660 mL/hr)-0.5mL, REF 383536; BD Q-Style Luer Access Split Septum BD Vialon Material; BD Instaflash Needle Technology; Rx Only, Sterile EO. BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon" catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, dual port adapter, vent plug and pre-attached BD Q-Syte Luer Access Split Septum. The needle and catheter are protected by a needle cover. A BD Q-Syte device or an end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. These devices have BD Instaflash" Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and dual port adapter are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green). BD Nexiva closed IV catheter systems are intended to be inserted into a patient s vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

Becton Dickinson Infusion Therapy Systems

Class I - Dangerous

There is a breach in the product packaging that renders the product non-sterile.

Sep 8, 2021 Diagnostic Equipment Nationwide View Details →

Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EJ STENT GRAFT ESBF2814C103EJ ENDUR IIS BIF ESBF2814C103EJ STENT GRAFT ESBF2814C103EJ ENDUR IIS BIF ESBF2814C103EJ STENT GRAFT ESBF2814C103EJ ENDUR IIS BIF ESBF2314C103E Stent Graft ESBF2314C103E Endur IIs BIF ESBF2314C103E Stent Graft ESBF2314C103E Endur IIs BIF ESBF2314C103E Stent Graft ESBF2314C103E Endur IIs BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2514C103EE STENT GRAFT ESBF2514C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2314C103EE STENT GRAFT ESBF2314C103EE ENDUR IIS BIF ESBF2314C103EE STENT GRAFT ESBF2314C103EE ENDUR IIS BIF ESBF2314C103EE STENT GRAFT ESBF2314C103EE ENDUR IIS BIF ESBF2314C103EE STENT GRAFT ESBF2314C103EE ENDUR IIS BIF ESBF2314C103EE STENT GRAFT ESBF2314C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2314C103EE STENT GRAFT ESBF2314C103EE ENDUR IIS BIF ESBF2314C103EE STENT GRAFT ESBF231

Medtronic Vascular

Class I - Dangerous

Due to, during stent graft deployment, the spindle may detach from the distal end of the spindle hypotube.

Oct 15, 2021 Implants & Prosthetics Nationwide View Details →

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Oct 20, 2021 Diagnostic Equipment Nationwide View Details →

Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

Sep 17, 2021 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Povidone-Iodine swabstick manufactured by PDI, Inc. recalled due to Out of Specifications was used as a component in some of the ROI CPS, LLC products.

Aug 10, 2021 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Povidone-Iodine swabstick manufactured by PDI, Inc. recalled due to Out of Specifications was used as a component in some of the ROI CPS, LLC products.

Aug 10, 2021 Other Medical Devices Nationwide View Details →

Vit Kit-Freeze is intended for use in assisted reproductive procedures for vitrification and storage of human oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. This kit is designed for use with CryoTip (Catalog #40709), and Vitrification Thaw Kit (Vit Kit-Thaw) for optimal recovery of specimens. Vitrification Freeze kit (which includes the VS Vitrification Solution) are cryopreservation media intended for use in the vitrification of oocytes (MII) and pronuclear (PN) zygotes, cleavage stage embryos, and blastocyst stage embryos. Catalog numbers: 90133-SO Vitrification Freeze kit, contained 9x1 mL. Vitrification solution, VS, Catalog #90132. Product form: Liquid. Equilibration Solution-ES is a HEPES buffered solution of Medium-199 containing gentamicin sulfate, 7.5% (v/v) of each DMSO and ethylene glycol and 20% (v/v) Dextran Serum Supplement (DSS). Vitrification Solution-VS is a HEPES buffered solution of Medium-199 containing gentamicin sulfate, 15% (v/v) of each DMSO and ethylene glycol, 20% (v/v) DSS and 0.5 M sucrose. DSS is a protein supplement consisting of 50 mg/mL therapeutic grade Human Serum Albumin (HSA) and 20 mg/mL Dextran. DSS is used at 20% (v/v) in Vit Kit-Freeze for a final concentration of 10 mg/mL HSA and 4 mg/mL Dextran. These two solutions are to be used in sequence according to the step-wise microdrop vitrification protocol. Vitrification Solution-VS Aseptically filtered. Test results reported on C of A Vit Kit -Freeze (Vitrification Freeze Kit) with Gentamycin and DSS Contains 2 x 1 mL Equilibration Solution- ES; 2 X 1 mL Vitrification Solution-VS ...90133-DCOS...RX Only...Sterile ...20YY-MM-DD

Fujifilm Irvine Scientific

Class I - Dangerous

Due to a component in kit being labeled with the incorrect Expiration Date

Sep 8, 2021 Diagnostic Equipment Nationwide View Details →

A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinine levels up to 9,000 mg/L.

Oct 18, 2021 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Povidone-Iodine swabstick manufactured by PDI, Inc. recalled due to Out of Specifications was used as a component in some of the ROI CPS, LLC products.

Aug 10, 2021 Other Medical Devices Nationwide View Details →

Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.

Jun 17, 2021 Other Medical Devices View Details →

Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.

Oct 8, 2021 Surgical Instruments Nationwide View Details →