ðŸĨ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

Nov 1, 2022 Surgical Instruments Nationwide View Details →

Device exceeds the limitations of a Class I 510(k)-exempt device so the user manual, instructions for use, and promotional materials are being corrected to remove all references to downstream applications.

Nov 2, 2022 Diagnostic Equipment View Details →

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Jan 27, 2022 Implants & Prosthetics Nationwide View Details →

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Oct 20, 2022 Infusion Pumps Nationwide View Details →
Class I - Dangerous

Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.

Oct 17, 2022 Diagnostic Equipment Nationwide View Details →

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.

Oct 25, 2022 Diagnostic Equipment View Details →

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

Oct 5, 2021 Diagnostic Equipment View Details →
Class I - Dangerous

Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.

Oct 17, 2022 Diagnostic Equipment Nationwide View Details →

Device is mislabeled. The package insert for this ASR device contains a chromomap - an image showing the position the FISH probe maps to on the specific chromosome. For this device the chromomap in version 1 of the package insert is incorrect and doesn't not match the probe design or specification contained on the package insert

Oct 27, 2022 Other Medical Devices View Details →
Class I - Dangerous

Due to difficulty removing the liner on the adhesive components of the affected drapes and as a result may render the product unusable. In addition, the manufacturer of the drapes has observed an increase in reported adhesive related skin injuries for these affected lots.

Oct 31, 2022 Surgical Instruments Nationwide View Details →

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

Oct 5, 2021 Other Medical Devices View Details →

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.

Oct 25, 2022 Diagnostic Equipment View Details →