Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
MAGNETOM Prisma. Model Number: 10849582.
Siemens Medical Solutions USA
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH
Medline Industries
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
BIOGRAPH One (DE). Model Number: 11689172.
Siemens Medical Solutions USA
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.
Mislabeling. The ELAN 4 FIXED DURAGUARD STANDARD was labeled as "Long" and the ELAN 4 FIXED DURAGUARD LONG was labeled as "Standard."
MAGNETOM Verio Dot. Model Number: 10684333.
Siemens Medical Solutions USA
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
Tandem Diabetes Care
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Allura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Allura Xper FD10/10 Product Codes: (1)722011, (2) 722027, (3) 722005; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.
Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Allura Xper FD10 OR Table Product Codes: (1)722022, (2) 722033; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
Stryker Communications
Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
Philips CT systems labeled as the following: 1. Brilliance iCT, Model Number: 728306.
Philips North America
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.