There is a potential that the distal end of the catheter may be bent beyond its intended curvature and when connected to patient monitoring device, the monitor may display an error message.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
BioPlex 2200, Anti-CCP Reagent Pack, REF 665-3250, IVD
Bio-Rad Laboratories
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.
Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
Howmedica Osteonics
Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.
Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure
EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108
Encore Medical
Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.
Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
A service unit was potentially improperly tested and calibrated during manufacturing on a failed test station.
Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE
Terumo Cardiovascular Systems
Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications
VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System, Model Number C-VH-4000
Maquet Cardiovascular
Some batches of product were not sterilized to their minimum sterilization specification.
Medtronic Grafton DBF Inject: a) GRAFTON T50303 DBF INJECT 3CC, REF T50303; b) GRAFTON T50306 DBF INJECT 6CC, REF T50306; c) GRAFTON T50309 DBF INJECT 9CC, REF T50309; Demineralized Bone Matrix (DBM) product
Medtronic Sofamor Danek USA
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
FreeStyle Libre Reader, REF: 71951-01, 71952-01, 71953-01 a component of the FreeStyle Libre 2 Flash Glucose Monitoring System
Abbott Diabetes Care
Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications
Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060
CooperSurgical
It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development.