ðŸĨ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly.

Nov 7, 2023 Diagnostic Equipment Nationwide View Details →

Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.

Nov 17, 2023 Diagnostic Equipment Nationwide View Details →

During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.

Oct 10, 2023 Other Medical Devices Nationwide View Details →

STERILE BETADINE, REF DYNDA1998

MEDLINE INDUSTRIES, LP - Northfield

Class I - Dangerous

The product description on the labeling includes the word "Sterile" or STRL", however the contents of the solution inside the foil sachet is non-sterile.

Nov 3, 2023 Other Medical Devices Nationwide View Details →

During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.

Oct 10, 2023 Other Medical Devices Nationwide View Details →

Certain factory settings for the device were incorrectly described in the IFU-related to the warning limits for PVen and PAux, and the default settings for the venous bubble sensor (VBS) and the automatic lock in the Minimized Extracorporeal Circulation (MECC) ThApp

Nov 6, 2023 Infusion Pumps Nationwide View Details →

STERILE PVP, REF DYNDA1649

MEDLINE INDUSTRIES, LP - Northfield

Class I - Dangerous

The product description on the labeling includes the word "Sterile" or STRL", however the contents of the solution inside the foil sachet is non-sterile.

Nov 3, 2023 Other Medical Devices Nationwide View Details →

Mathematical modeling of the 4Kscore with and without DRE information was recently performed on over 3,000 specimens. DRE information is a required component of The 4Kscore Test.

Oct 31, 2023 Diagnostic Equipment Nationwide View Details →

GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly.

Nov 7, 2023 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.

Jul 13, 2020 Other Medical Devices View Details →