ðŸĨ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to extended surgery time

Nov 30, 2023 Implants & Prosthetics Nationwide View Details →

It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

Nov 6, 2023 Surgical Instruments Nationwide View Details →

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Nov 10, 2023 Other Medical Devices Nationwide View Details →

Tubes have missing additive/anticoagulant resulting in clotting in most cases. A completely clotted sample would lead to no results being reported, which could lead to a delay in treatment or diagnosis. An erroneous glucose or lactose result from a partially clotted sample could lead to inappropriate treatment.

Dec 12, 2023 Other Medical Devices Nationwide View Details →

If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.

Dec 7, 2023 Infusion Pumps Nationwide View Details →

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Nov 10, 2023 Other Medical Devices Nationwide View Details →

Maquet Equipment OR Light Systems, Model Numbers and Catalog Numbers BRAINLAB CAMERA - ARDEQT269000A, BRAINLAB CAMERA - ARDEQT269000C, BRAINLAB CAMERA - ARDEQT269001A, BRAINLAB CAMERA - ARDEQT269001C, BRAINLAB CAMERA - ARDEQT269004A, BRAINLAB CAMERA - ARDEQT269004C, BRAINLAB CAMERA - ARDEQT269005A, BRAINLAB CAMERA - ARDEQT269005C, EQTBRB SAT12 - ARDSAT269000A, EQTBRB SAT13 - ARDSAT269001A, EQTFHS010 11 - ARDEQT239014A, EQTFHS010 SAT12 - ARDSAT239005A, EQTFHS010 SAT13 - ARDSAT239006A, EQTFHS010 SAT16 - ARDSAT239023A, EQTMHD231 - ARDEQT239013A, EQTMHD231 11 - ARDEQT239016A, EQTMHD231 16 - ARDEQT239023A, EQTMHD231 SAT12 - ARDSAT239013A, EQTMHD231 SAT13 - ARDSAT239014A, EQTMHS013 08 - ARD2EQT00010A, EQTMHS013 11 - ARDEQT239007A, EQTMHS013 14 - ARDEQT239008A, EQTMHS013 16 - ARDEQT239009A, EQTMHS013 SAT12 - ARDSAT239007A, EQTMHS013 SAT13 - ARDSAT239008A, EQTMHS021 11 - ARDEQT239010A, EQTMHS021 14 - ARDEQT239011A, EQTMHS021 16 - ARDEQT239012A, EQTMHS021 SAT12 - ARDSAT239009A, EQTMHS021 SAT13 - ARDSAT239010A, EQTSC07 SAT12 - ARD515031140A, EQTSC07 SAT13 - ARD515032140A, EQTSC07 SAT16 - ARD2SAT00120A, EQTSC07 SAT16 - ARDSAT229007A, EQTSHD 11 - ARDEQT229001A, EQTSHD SAT - ARD515031990A, EQTSHD SAT - ARD515032990A, EQTSPC12 08 - ARDEQT269008A, EQTSPC12 11 - ARDEQT269009A, EQTSPC12 SAT12 - ARDSAT269005A, EQTSPC12 SAT13 - ARDSAT269006A, EQTXHD127 08 - ARDEQT239026A, EQTXHD127 11 - ARDEQT239027A, EQTXHD127 16 - ARDEQT239028A, EQTXHD127 SAT12 - ARDSAT239021A, EQTXHD127 SAT13 - ARDSAT239022A, EQTXHS010 08 - ARD2EQT00030A, EQTXHS010 08 - ARDEQT239024A, EQTXHS010 10 - ARD2EQT00050A, EQTXHS010 11 - ARDEQT239017A, EQTXHS010 14 - ARDEQT239019A, EQTXHS010 16 - ARDEQT239020A, EQTXHS010 SAT12 - ARDSAT239015A, EQTXHS010 SAT13 - ARDSAT239018A, EQTXHS021 08 - ARDEQT239025A, EQTXHS021 11 - ARDEQT239018A, EQTXHS021 14 - ARDEQT239021A, EQTXHS021 16 - ARDEQT239022A, EQTXHS021 SAT12 - ARDSAT239019A, EQTXHS021 SAT13 - ARDSAT239020A, EQTXS3216 11 - ARDEQT239015A, EQTXS3216 SAT12 - ARDSAT239016A, EQTXS3216 SAT13 - ARDSAT239017A, HOR 14 XD2 - ARDEQT239006A, HOR 14 XD2 - ARDEQT239006C, HOR 14 XO - ARDEQT269002A, HOR 14 XO - ARDEQT269002C, HOR 14 XS0 - ARDEQT239004A, HOR 14 XS0 - ARDEQT239004C, HOR 14 XS1 - ARDEQT239005A, HOR 14 XS1 - ARDEQT239005C, HOR 16 XD2 - ARDEQT239003A, HOR 16 XD2 - ARDEQT239003C, HOR 16 XO - ARD267900080A, HOR 16 XO - ARDEQT269003A, HOR 16 XO - ARDEQT269003C, HOR 16 XS0 - ARDEQT239001A, HOR 16 XS0 - ARDEQT239001C, HOR 16 XS1 - ARDEQT239002A, HOR 16 XS1 - ARDEQT239002C, SAT BRAINLAB 12 - ARD515070110A, SAT BRAINLAB 12 - ARD515070110C, SAT BRAINLAB 13 - ARD515071110A, SAT BRAINLAB 13 - ARD515071110C, SAT BRAINLAB CAMERA - ARDSAT269000C, SAT BRAINLAB CAMERA - ARDSAT269001C, SAT BRAINLAB CAMERA - ARDSAT269003A, SAT BRAINLAB CAMERA - ARDSAT269003C, SAT BRAINLAB CAMERA - ARDSAT269004A, SAT BRAINLAB CAMERA - ARDSAT269004C, SAT CAM 12 - ARD515031880A, SAT CAM 12 - ARD515031880C, SAT CAM 13 - ARD515032880A, SAT CAM 13 - ARD515032880C, SAT MEDIA - ARD2SAT00210A, SAT MEDIA - ARD2SAT00220A, SAT MHD214 - ARDSAT239011A, SAT MHD214 - ARDSAT239012A, SAT ORCHID - ARDSAT229000A, SAT ORCHID - ARDSAT229000C, SAT ORCHID - ARDSAT229001A, SAT ORCHID - ARDSAT229001C, SAT ORCHID - ARDSAT229003A, SAT ORCHID - ARDSAT229003C, SAT ORCHID - ARDSAT229004A, SAT ORCHID - ARDSAT229004C, SAT ORCHIDE 12 - ARD515031991A, SAT ORCHIDE 12 - ARD515031991C, SAT ORCHIDE 12 - ARDSAT229005A, SAT ORCHIDE 12 - ARDSAT229005C, SAT ORCHIDE 13 - ARD515032991A, SAT ORCHIDE 13 - ARD515032991C, SAT ORCHIDE 13 - ARDSAT229006A, SAT ORCHIDE 13 - ARDSAT229006C, SAT PERF 258S - ARD515053130A, SAT PERF 258S - ARD515053130C, SAT SC 071 - ARD515031140C, SAT SC 071 - ARD515032140C, SAT SC07 HOR - ARD2SAT00120C, SAT SHD 12 - ARD515031990C, SAT SHD 13 - ARD515032990C, SAT SHD HOR - ARD2SAT00150A, SAT SHD HOR - ARD2SAT00150C, SAT SSD 13 - ARD515032080A, SAT SUP CAM 12 - ARD515031730A, SAT SUP CAM 13 - ARD515032730A, SAT XD 241 - ARD515031660A, SAT XD 241 - ARD515031660C, SAT XD 241 - ARD515032660A, SAT X

Getinge Usa Sales

Class I - Dangerous

A potential for a light system to fall in the operating room.

Nov 8, 2023 Other Medical Devices Nationwide View Details →

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF - ARDVCS209011A, VCS40SF 08 - ARDVCS209046A, VCS44DF - ARDVCS209005A, VCS44DF - ARDVCS209005C, VCS44DF - ARDVCS209016A, VCS44SF - ARDVCS209004A, VCS44SF - ARDVCS209004C, VCS44SF - ARDVCS209015A, VCS44SFDF - ARD2VCS00040A, VCS60DF - ARDVCS209003A, VCS60DF - ARDVCS209003C, VCS60DF - ARDVCS209014A, VCS60DF - ARDVCS209022A, VCS60DF 11 - ARDVCS209047A, VCS60SF - ARDVCS209002A, VCS60SF - ARDVCS209002C, VCS60SF - ARDVCS209013A, VCS64DF - ARDVCS209007A, VCS64DF - ARDVCS209007C, VCS64DF - ARDVCS209018A, VCS64SF - ARDVCS209006A, VCS64SF - ARDVCS209006C, VCS64SF - ARDVCS209017A, VCS64SFDF - ARD2VCS00010A, VCS64SFDF 1011 - ARDVCS209048A, VCS66DF - ARDVCS209009A, VCS66DF - ARDVCS209009C, VCS66DF - ARDVCS209020A, VCS66SF - ARDVCS209008A, VCS66SF - ARDVCS209008C, VCS66SF - ARDVCS209019A, VCSII40DF SB08 - ARDVCS209023A, VCSII40DF SB11 - ARDVCS209024A, VCSII40SF SB11 - ARDVCS209025A, VCSII44DF SB0810 - ARDVCS209026A, VCSII44SF SB1011 - ARDVCS209027A, VCSII60DF SB08 - ARDVCS209028A, VCSII60SF SB11 - ARDVCS209029A, VCSII64DF SB0810 - ARDVCS209030A, VCSII64SF SB1011 - ARDVCS209031A, VCSII66DF SB0810 - ARDVCS209032A, VCSII66SF SB1011 - ARDVCS209033A, VCSII40DF TK SB08 - ARDVCS209034A, VCSII40DF TK SB11 - ARDVCS209035A, VCSII40SF TK SB11 - ARDVCS209036A, VCSII44DF TK SB0810 - ARDVCS209037A, VCSII44SF TK SB1011 - ARDVCS209038A, VCSII60DF TK SB08 - ARDVCS209039A, VCSII60SF TK SB11 - ARDVCS209040A, VCSII64DF TK SB0810 - ARDVCS209041A, VCSII64SF TK SB1011 - ARDVCS209042A, VCSII66DF TK SB0810 - ARDVCS209043A, VCSII66SF TK SB1011 - ARDVCS209044A, VCS40DF SAT 12 - ARD2VCS00070A, VCSII40SF SB08 - ARDVCS209049A, VCSII64SFDF SB1011 - ARDVCS209050A, SAT VLT400D - ARD2SAT00080C, SAT VLT400D - ARD2SAT00100C, SAT VLT400DF - ARD2SAT00080A, SAT VLT400DF - ARD2SAT00100A, SAT VLT400S - ARD2SAT00030C, SAT VLT400S - ARD2SAT00040C, SAT VLT400SF - ARD2SAT00040A, SAT VLT40DF - ARD268800130A, SAT VLT40DF - ARD268800130C, SAT VLT40SF - ARD268800120A, SAT VLT40SF - ARD268800120C, SAT VLT60DF - ARD2SAT00170A, SAT VLT60DF - ARD2SAT00170C, SAT VLT60SF - ARD2SAT00180C, SAT VST40DF - ARDSAT209018A, SAT VST40DF - ARDSAT209019A, SAT VST40DF - ARDSAT209022A, SAT VST40DF - ARDSAT209023A, SAT VST40SF - ARDSAT209010A, SAT VST40SF - ARDSAT209011A, SAT VST40SF - ARDSAT209014A, SAT VST40SF - ARDSAT209015A, SAT VST60DF - ARDSAT209034A, SAT VST60DF - ARDSAT209035A, SAT VST60DF - ARDSAT209038A, SAT VST60DF - ARDSAT209039A, SAT VST60SF - ARDSAT209026A, SAT VST60SF - ARDSAT209027A, SAT VST60SF - ARDSAT209030A, SAT VST60SF - ARDSAT209031A, STANDOP 40DF SINGLE - ARDSTP229005A, STANDOP 40SF SINGLE - ARDSTP229000A, STANDOP 44DF DOUBLE - ARDSTP229007A, STANDOP 44SF DOUBLE - ARDSTP229002A, STANDOP 60DF SINGLE - ARDSTP229006A, STANDOP 60SF SINGLE - ARDSTP229001A, STANDOP 64DF DOUBLE - ARDSTP229008A, STANDOP 64SF DOUBLE - ARDS

Getinge Usa Sales

Class I - Dangerous

A potential for a light system to fall in the operating room.

Nov 8, 2023 Other Medical Devices Nationwide View Details →

Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.

Dec 13, 2023 Other Medical Devices Nationwide View Details →
Class I - Dangerous

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

Nov 20, 2023 Other Medical Devices Nationwide View Details →

A potential for a light system to fall in the operating room.

Nov 8, 2023 Other Medical Devices Nationwide View Details →

A potential for a light system to fall in the operating room.

Nov 8, 2023 Other Medical Devices Nationwide View Details →