Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
2008T HD SYS. CDX BLUESTAR
Fresenius Medical Care Holdings
Potential PCBA leaching from tubing of hemodialysis machines.
Venue Go Standard Carts Ref: (H45181VC and H45103VCW), used with Venue Go R2, R3, R4 systems R2, R3, and R4 systems; pulsed doppler imaging system
GE MEDICAL SYSTEMS, ISRAEL
Some Venue Go Standard Carts can develop an internal failure of the vertical/horizontal tilt adjustment mechanism which can result in the Venue Go system detaching from the cart and falling.
Guider Softip Guide Catheter 5F 100CM MULTI PURPOSE, Catalog number M003101630, cardiac guide catheter
Boston Scientific
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Guider Softip Guide Catheter 8F 90CM MULTI PURPOSE, Catalog number H965100480, cardiac guide catheter
Boston Scientific
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle and the rare events of retinal vasculitis
This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.
Guider Softip Guide Catheter 5F PRE-SHAPED 40 90CM, Catalog number M003100620, cardiac guide catheter
Boston Scientific
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
Medical convenience kit contains Zimmer Dermatome Blades which were recalled by Zimmer due to 38 complaints received related to skin grafts being thin and non-uniform when using the affected blades.
Incisive CT, software version 5.0
Philips Healthcare
Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Guider Softip Guide Catheter 6F 90CM MULTI PURPOSE, Catalog number H965100460, cardiac guide catheter
Boston Scientific
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Guider Softip Guide Catheter 6F PRE-SHAPED 40 90CM, Catalog number H965100420, cardiac guide catheter
Boston Scientific
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO
Becton Dickinson Infusion Therapy Systems
There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
EarlyVue VS30 Vitals Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863380
Philips North America
A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Centurion Sterile CirClamp with 1.45cm Bell Reprocessed, Reorder CR3145K, 1 per package, 12 per case.
MEDLINE INDUSTRIES, LP - Northfield
The kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size.