Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Intera 1.5T Power/Pulsar
Philips North America
Potential component failure in the Gradient Coil could product smoke and/or fire.
Intera 1.5T
Philips North America
Potential component failure in the Gradient Coil could product smoke and/or fire.
Due to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
VariSoft infusion set, Model Number 1002830
Unomedical A/S
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
Alinity c Iron Reagent, Reference Number 08P3920
Sentinel CH SpA
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Glucose and Lactate, Part Numbers 00024315009, 00024330009, 00024345009, 00024360009
Instrumentation Laboratory
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
Model 97745 Controller used with Models 97715 and 97716 Intellis Neurostimulators
Medtronic Neuromodulation
Units distributed in Korea and Turkey that did not have the correct firmware installed to support the user interface in the local language.
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Certain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not being made with Di(2-ethylhexyl) phthalate (DEHP); however, Avanos has determined that the tubing components of the product did contain DEHP.