RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
Three software issues affecting incorrect image display, error interpreting patient images due to image reporting, and patient exposure to incorrect image/function during clinical use.
Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.
Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
One product complaint was received reporting that there was an incorrect adapter in the packaging. An adapter with neck length M/+0 was inside the product packaging, which should have contained an adapter with neck length S/-3.0. The difference in neck length may be recognized by the size (S or M) indicator on the device.
The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
Sodium chloride irrigation USP, and sterile water for irrigation USP, that may be packaged inside kits, have the potential for a lack of sterility assurance, which could result in a nonsterile product, use of which could cause infection.
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.