ðŸĨ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

Specific lots of Neria Soft Infusion Sets, which have been shipped into the US Market, have been shipped with the incorrect Instruction for Use (IFU) within the final pack. The IFU included is the IFU registered for the Rest of World (ROW) countries. The difference between the US IFU and the ROW IFU is the addition of an intended use section in the ROW IFU.

Mar 5, 2024 Infusion Pumps Nationwide View Details →

Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that have not been flushed out over time with usage. Exposure to PCBAs could cause: skin conditions, liver damage, impaired reproduction, development of certain kinds of cancer in humans.

Mar 6, 2024 Other Medical Devices Nationwide View Details →

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

Dec 19, 2023 Surgical Instruments View Details →