icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Model No. CYF-VH, CYSTO-NEPHRO VIDEOSCOPE
Shirakawa Olympus Co.
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
BrightView X, Gamma Camera, Product Code 882478
Philips North America
While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
(1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)
Philips North America
The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)
CORI REAL INTELLIGENCE Robotic Drill, Product No. ROB10013 and ROB10013S
Blue Belt Technologies
Possibility that the user is unable to resolve drill disconnection error messages.
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Medtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUS
Medtronic Navigation
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Model No. CYF-VHR, CYSTO-NEPHRO VIDEOSCOPE
Shirakawa Olympus Co.
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Baxter FloSeal Hemostatic Matrix NL, 5 mL, Product Codes: ADS201883
Baxter Healthcare
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
New User Manual with additional instructions for use and a hose and mask adapter provided by the device manufacturer. This voluntary recall is intended to help reduce potential health risks of exposure to ozone gas after the use of a SoClean device.
Cordis became aware that the label for the carton box containing 10 units is incorrect for some boxes of two lots of MynxGrip. The carton box from lot F2322902 (Item MX5021, MynxGrip 5F) as labeled as lot F2322903 (MX6721, MynxGrip 6/7F) and product from lot F2322903 (MX6721, MynxGrip 6/7F) was labeled as F2322902 (Item MX5021, MynxGrip 5F). The primary label on individual devices is correct.
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Medtronic, REF 9733236, Pin, 150mm, STERILE, PERCUTANEOUS
Medtronic Navigation
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
Model No. BF-XP190, EVIS EXERA III Bronchovideoscope
Shirakawa Olympus Co.
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.