New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
The microcatheter IFU contains an intended use that has not been evaluated by the FDA.
ProxiDiagnost upgrade- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref:706150
Philips North America
Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)
Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures. Models: 5430-30151, 5430-30152, and 5430-30154
Mirion Technologies (Capintec)
Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator
One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the correct expiration date for the pouches is September 12, 2024. The kits are labeled with an expiration date of April 30, 2025, however, the correct expiration date for the kits is September 12, 2024. The issue was discovered on November 28, 2023 by a supply chain employee who determined that the kits had expiration dates beyond the FDA authorized shelf life.
Single Use Distal Cover MAJ-2315 Used with: EVIS EXERA III Duodenovideoscope TJF-Q190V- designed to be attached to Olympus duodenovideoscopes to cover the distal end of the insertion tube and around the forceps elevator
Olympus Corporation of the Americas
Single Use Distal Cover MAJ-2315 may unexpectedly detach from the duodenoscope during use result include the risk of mucosal injury, tissue damage, bleeding, perforation, or obstruction due to retained distal cover in the gastrointestinal tract that would require urgent medical intervention for removal, and/or burns from the use an uncovered duodenoscope distal end. Detachment of a distal cover in the oral cavity may result in aspiration, airway obstruction, or respiratory distress and would require immediate medical intervention for removal.
ProxiDiagnost N90- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref: (1)706100 (2)706110
Philips North America
Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
3:1 Dermacarrier, Model Number 00219501300, skin graft carrier
Zimmer Surgical
Due to manufacturing issue, the product is labeled as 3:1 ratio, but the ridge pattern is a 1.5:1 ratio. The problem may lead to minor tissue damage or additional unplanned graft if deemed necessary.
CombiDiagnost R90 1.1 (709031)
PHILIPS MEDICAL SYSTEMS
While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with the CARDIOHELP heart-lung machine.
Maquet Medical Systems USA
The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.
GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500
Wipro GE Healthcare Private
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
Mobile x-ray systems moving arm frame welding issue which presents the potential of bodily harm due to fall of the arm.
GE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646, M1166822
Wipro GE Healthcare Private
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
AFT (Allograft Filler Tube) Straight Tube (1/2 Filled), Product Code 227010 (bone void filler)
Musculoskeletal Transplant Foundation
Filling process qualification did not demonstrate adequate process capability. Possible consequences include extended surgery time due to extrudability issues.
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850, 6600-0081-850, 6600-0961-500, 6600-0083-850
Wipro GE Healthcare Private
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
AFT (Allograft Filler Tube) Diverted Tube (3/4 Filled), Product Code 227005 (bone void filler)
Musculoskeletal Transplant Foundation
Filling process qualification did not demonstrate adequate process capability. Possible consequences include extended surgery time due to extrudability issues.