The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy
The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
The wrong configuration of the blade was in the package. The bevel was on the opposite side, instead of being in front (bevel up) it was in the back (bevel down).
Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature monitoring devices.
The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.
The wrong configuration of the blade was in the package. The bevel was on the opposite side, instead of being in front (bevel up) it was in the back (bevel down).
A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.
Exactech is recalling all affected UHMWPE (ultra-high molecular weight polyethylene) knee patella components packaged in out of specification vacuum bags.
Exactech is recalling all affected UHMWPE (ultra-high molecular weight polyethylene) knee patella components packaged in out of specification vacuum bags.
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.