Expired Products distributed to customers
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS Diluent Pack 3 comes with 3 packs per sales unit.
Ortho-Clinical Diagnostics
VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.
WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-01
WOM World Of Medicine AG
Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery
The reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000 Access Immunoassay Analyzer may generate a System Event log message that indicates PnP errors. Any cancelled tests will have an associated SYS flag. The analyzer may enter the Red-system status because of this issue. The analyzer also may not eject RVs, where unbound conjugate could potentially splash up on the sidewall of the RV, not being properly washed away, resulting in an artificially higher signal (RLU). The analyzer may not detect this splashing and continue with normal operation, which can cause false low or high erroneous results. Consequently, the analyzer may cancel in-progress tests. These errors could potentially cause delay reporting patient test results and patient treatments. The degree of the potential health hazard is dependent on the analytes that are potentially delayed, especially those analytes that would be used in the acute setting when treatment decisions are immediate. In this particular recall the highest risk analyte is troponin. Health consequences could include permanent impairment or possibly death if medical treatment decisions are delayed due to a delay in obtaining patient results. The issue was identified by Beckman Coulter Inc. manufacturing staff at their location in Brea California on February 29, 2024, during in-process acceptance inspection activities.
HORIBA custom configured fluorescence instrument, modular Fluorolog-QM
Horiba Instruments Incorporated
HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Siemens AG/Siemens Healthcare GmbH
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
Sterile surgical convenience kit: Medline Retrograde PK-22025203-LF, Pack #DYNJ47873L, containing a Pure Pouch with component Sterile Curved Kelly Hemostat, Reorder #66175.
MEDLINE INDUSTRIES, LP - Northfield
Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.
To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.
First Source iQ Flex M. Portable cart and platform that supports an X-ray system and laptop computer.
First Source
Potential for the bolt that secures the positioning arm to the gas spring to come loose and fall onto patient.
Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).
GE Medical Systems Information Technologies
Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.
Avalon Ultrasound Transducer. Directs low-energy ultrasound beam toward the fetal heart and externally monitors fetal heart rate.
Philips North America
Potential for inaccurate fetal heart rate measurements when monitoring multiples.
EVOLUTIONΒΏ MP Femoral CS/CR Non-Porous Non-porous MicroPort total knee replacement implants are for cemented use only.
MicroPort Orthopedics
One confirmed incident has been received that EFSRN4PR, lot number MP1982170, Size 4 Evolution MP Non-Porous Femoral Right was opened during surgery and contained part number EFSRN3PL, lot number MP1982097, Size 3 Evolution MP Non-Porous Femoral Left in the packaging.
Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.
Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.
COBALT HF QUAD CRT-D MRI SureScan, Model Number DTPB2QQ, Implantable Cardioverter Defibrillator
Medtronic
Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.
Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.
Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.
Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.
Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support System
Thoratec Switzerland GMBH
Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.