Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
Firm officials reported to CIN-DO that their Electrosurgical Monopolar Reusable Bayonet Forceps are misbranded in that the label states that the tip size is: (0.5mm) when the tip size actually measures (1.0mm).
i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540
Abbott Point Of Care
Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
HOTLINE 3 Blood and Fluid Warmer( Reorder No. HL-390 ) disposable is not correctly installed , the microswitch may become damaged and unusable
FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.
Maquet Cardiovascular Us Sales
A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.
Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports.
Stryker Orthopaedics has become aware of the potential for cracking of the PMMA coating on their PM COCR Hip Stem product.
Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu
The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.
Firm officials reported to CIN-DO that the outer carton label of this sterile, single use device is misbranded in that it lists an incorrect expiration date. The outer carton label lists the sterile expiration date as: 2012-02-(28) [the manufacture date] ; however the device pouch label lists the sterile date as: 2017-02-(28). The pouch label lists the correct sterile expiration date.
Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.
Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. Occlusion malfunctions include stiff/difficult to adjust occlusion, loss of occlusion setting, unequal roller-to-roller occlusion and inability to adjust the occlusion. If it is
Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.
Drager PT 4000 Phototherapy System
Draeger Medical Systems
A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b
Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.
Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.