Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model TS057HF7. For use in patients who require hemodynamic monitoring.
Edwards Lifesciences
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 131HF7J. For use in patients who require hemodynamic monitoring.
Edwards Lifesciences
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 141HF7J. For use in patients who require hemodynamic monitoring.
Edwards Lifesciences
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Zimmer investigation following a complaint found one lot of NCB Polyaxial Locking Plate System NCB screws, 26mm, distributed in the United States contained incorrect 32 mm screws.
Content entered into the "Conclusions" free text box on the Report Writer screen was not represented on the final printed report.
BrainPath sheaths were mislabeled. Sheaths which measure 50mm were labeled as 60mm. The length callout on the label does not match the actual length of the sheath within the packaging resulting in an improper fit as further explained below. This mislabeling was discovered on 29-June-2012 during a clinical case.
Ortho Clinical Diagnostics revised the VITROS Chemistry Products DGXN Slides Instructions for Use (IFU) and removed heparin plasma as a recommended specimen type for VITROS DGXN Slides
Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.
Philips Healthcare
Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 141HF7P. For use in patients who require hemodynamic monitoring.
Edwards Lifesciences
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Sunquest Laboratory SMART & Sunquest Laboratory SMART Select
Sunquest Information Systems
In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur: 1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (<Y>/N)]=Y. 2) The Tests are ordered on the same accession number and 3) The Test codes are defined to have the same upload code as either replicate or
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 151F7. For use in patients who require hemodynamic monitoring.
Edwards Lifesciences
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stainle
In connection with a Warning Letter received from the U.S. Food & Drug Administration on December 9, 2011, DePuy Orthopaedics is taking steps to retrieve custom device implants that surgeons may have ordered from DePuy Orthopaedics, but not implanted. DePuy Orthopaedics received a Warning Letter from the FDA on December 9, 2011 regarding custom device implants. DePuy Orthopaedics has discontinue
The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with Alaris Pump Module (model 8100) manufactured between August 2010 and July 2011.
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 141F7. For use in patients who require hemodynamic monitoring.
Edwards Lifesciences
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
GE Healthcare is informing you of incorrect measurements when using the Aorta analysis protocol or customized protocols. Until a new software version is installed they are asking you not to use the aorta analysis protocol or the customized protocols for an aneurysm. A second issue was also discovered internally. When loading PET data into the Volume Viewer, if slices are missing in a PET serie
Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 131HF7P. For use in patients who require hemodynamic monitoring.
Edwards Lifesciences
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
3mL 9NC Coagulation Sodium Citrate 3.2%, Item #454334, Greiner Bio-One North America Inc., 4238 Capital Drive, Monroe, NC 28110. Intended Use: Venous blood collection tubes.
Greiner Bio-One North America
Part of the lot received a lower concentration of citrate solution which may cause falsely lower values for common coagulation tests.