๐Ÿฅ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or auto displayed, dismissed by the user, and then a different patient and signed exam is selected in single patient view. The Notes Viewer will

Oct 5, 2012 Infusion Pumps Nationwide View Details โ†’

ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the Air Column Detect Sensor halting the injection, and normally replacement of the BT2000 Automated Manifold Kit is necessary in order to proceed w

Mar 15, 2013 Infusion Pumps Nationwide View Details โ†’

The recall was initiated because Carefusion has identified potential risk associated with bolus programming with the Alaris PC unit (model 8015) software version 9.12. Automating infusion pump parameter input from either the electronic medical record (EMR) or the Alaris Auto-ID module for a continuous infusion with a bolus dose option may result in a loss of information from the Guardrails Data s

Mar 6, 2013 Diagnostic Equipment Nationwide View Details โ†’

As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co

Oct 21, 2011 Diagnostic Equipment Nationwide View Details โ†’

Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista(R) LOCI(R) Digoxin(DIGXN) Flex(R) reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL.

Feb 21, 2013 Diagnostic Equipment Nationwide View Details โ†’

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROSยฟ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.

Feb 7, 2013 Other Medical Devices Nationwide View Details โ†’

Various reciprocating and oscillating saw blades were incorrectly labeled. The labeling errors include: (1) The pouch says oscillating saw blade when it actually contains a reciprocating saw blade; (2) The box bar code date does not match the human readable expiration date; and (3) The label incorrectly states use with Zimmer/Synthes handset and not the Zimmer Universal Power System.

Apr 2, 2013 Implants & Prosthetics Nationwide View Details โ†’

Two unsealed pouches were found at an Edwards' distribution location in Japan. Incorrectly sealed pouches may result in a sterility barrier breach.

Mar 14, 2013 Other Medical Devices Nationwide View Details โ†’

The product has a Lithium Heparin contaminant which could adversely affect coagulation results potentially causing erroneous lab results.

Feb 6, 2013 Surgical Instruments Nationwide View Details โ†’

Depending on the system configuration, a software error message in versions 5.0.1 and 5.0.2. occurs if a beam plan is transmitted from XKnife using DICOM-RT.

Dec 10, 2012 Other Medical Devices View Details โ†’

During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.

Mar 21, 2013 Infusion Pumps Nationwide View Details โ†’

The device is being recalled because the firm has received multiple complaints related to incorrect assembly that may result in dilution of blood samples.

Jan 30, 2013 Diagnostic Equipment Nationwide View Details โ†’

Firm initiated an update to Directions for Use necessitating replacement of previously distributed Directions for Use.

Sep 7, 2012 Surgical Instruments Nationwide View Details โ†’

GE Healthcare conducted a recall on the Muse v7 because incorrect medications were displayed on the MUSE, which could result in incorrect treatment.

Jan 4, 2012 Other Medical Devices View Details โ†’