Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
MicroScan WalkAway-40 plus Instrument and MicroScan Walkaway-96 plus instrument access door hinge. Siemens Healthcare Diagnostics, Inc. in vitro diagnostic
Siemens Healthcare Diagnostics
Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers.
Connection QRG for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System . Intuitive Surgical Endoscopic Instrument Control System.
Intuitive Surgical
Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the Emergency Grip Release Wrench, part number 710142.
Zimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
Laceration/Suture Removal, Sterile Single Use, Latex Free, 20/case, Medical Action Industries, Inc, Arden, NC.
Medical Action Industries
Some of the sterile kits may have incomplete seals which might compromise the sterility of the kits.
CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.
Nihon Kohden America
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
Microline Surgical ReNew Laproscopic Fenestrated Grasper Forceps Tips (Reusable) Ref 3222
Microline Surgical
Potential for the Fenestrated Grasper Tip jaw to break when force is applied to the jaw
Stapler Addendum for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System. Intuitive Surgical Endoscopic Instrument Control System.
Intuitive Surgical
Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the Emergency Grip Release Wrench, part number 710142.
ADVIA CentaurΒΏ Systems TnI-Ultra assay; Catalog Number 02789602 (100 tests), SMN 10317708 and 02790309 (500 tests), SMN 10317790
Siemens Healthcare Diagnostics
The solid phase reagent in some of the ADVIA CentaurΒΏ TnI -Ultra ReadyPacksΒΏ, lot 084, is darker in appearance, creating a potential for incorrect control and patient results.
Laceration Trays, Sterile Single Use, Latex Free, 20/case, Medical Action Industries, Inc., Arden, NC.
Medical Action Industries
Some of the sterile kits may have incomplete seals which might compromise the sterility of the kits.
BD FACS Canto and associated Fluidics Cart Product Usage: Flow cytometer for use in In vitro Diagnostics.
BD Biosciences, Systems & Reagents
Laser cooling fans and internal fan of the associated fluidics cart are not operational.
Insert for IRK for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System. Intuitive Surgical Endoscopic Instrument Control System.
Intuitive Surgical
Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the Emergency Grip Release Wrench, part number 710142.
Vision Troubleshooting Guide for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System. Intuitive Surgical Endoscopic Instrument Control System.
Intuitive Surgical
Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the Emergency Grip Release Wrench, part number 710142.
Image taken by AFC-330 has a white spot which may affect diagnosis or evaluation of image.
If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.
Cautery QRG for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System . Intuitive Surgical Endoscopic Instrument Control System.
Intuitive Surgical
Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the Emergency Grip Release Wrench, part number 710142.
There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.
Vessel Sealer QRG for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System. Intuitive Surgical Endoscopic Instrument Control System.
Intuitive Surgical
Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the Emergency Grip Release Wrench, part number 710142.
ACUSON S Family Ultrasound Systems at software versions VC25D, VC30A and VC30B; (S2000, S3000 and S1000 systems) Siemens Medical Solutions USA, Inc. Ultrasound imaging systems.
Siemens Medical Solutions USA
There is a potential measurement error with the Auto-Stats (auto statistics) measurement tool on the ACUSON S Family ultrasound system.
Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.