Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.

Jul 16, 2014 Diagnostic Equipment Nationwide View Details →

Some customers reported that while using the DX-D600 in room or exam preparation, the Overhead Tube Crane (OTC) showed increased velocity and uncontrolled longitudinal movement.

Jul 2, 2014 Diagnostic Equipment View Details →

The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction (EF) displayed should be the same as the ejection fraction (EF) originally displayed when the bookmark was first created. In some instances, the ejection fraction (

Jul 27, 2014 Diagnostic Equipment Nationwide View Details →

The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring.

Jul 23, 2014 Other Medical Devices Nationwide View Details →

To clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).

Jul 3, 2014 Implants & Prosthetics Nationwide View Details →

Integra LifeSciences is recalling the Integra NewPort MIS System because there is a possibility that components from the demonstration set may have been mixed with the NewPort MIS pedicle screw system components used in surgery.

Jun 27, 2014 Implants & Prosthetics View Details →

Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.

Apr 17, 2013 Diagnostic Equipment Nationwide View Details →

To clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).

Jul 3, 2014 Implants & Prosthetics Nationwide View Details →

Manufacturer made a change in the production process that resulted in a change in final gel weight. RECALL EXPANDED 7/8/2014 Firm expanded their recall to include all lots of product.

Aug 7, 2013 Surgical Instruments Nationwide View Details →

Zimmer is initiating a voluntary recall of 20 production lots of Moore Hip Prosthesis Long Fenestrated Stems due to packaging design verification test failures. Specifically, 9 of 16 test samples failed simulated distribution and shipping testing when the Moore hip stem breached either the inner or outer package tray. Zimmer has not received any complaints of damaged package trays for these 20 lot

Jul 1, 2014 Implants & Prosthetics Nationwide View Details →

Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne

Jul 11, 2014 Implants & Prosthetics Nationwide View Details →