Dako is recalling the Dako Link 4.0.3 software on the Dako Autostainer Link 48 because Dako has determined that an upgrade to this version of the software may cause premature rejection of reagent bottles or inadequate aspiration of reagent if the instrument has also had a preventative maintenance or waste pump replacement, and the baseplate is not completely aligned.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).
No packaging. The device is a computed tomography scanner. Intended to provide corss sectional images of the human body and visualization of the internal organs of the body.
Philips Medical Systems (Cleveland)
When the Extended Display Firled ov View reconstruction is used, the CT numbers and anatomy in teh extended region are not accurate.
UltraStream Chronic Hemodialysis Catheter may develop cracks on the red (arterial) luer during attachment or manipulation.
ICU is recalling the ConMed Stat 2 flow controller because it may deliver a significantly higher flow rate than was intended.
Disposable Patient Circuit that allows water to leak into the center gas lumen
Disposable Patient Circuit that allows water to leak into the center gas lumen
The CareFusion AirLife Heated Infant Breathing Circuit is being recalled due to a regulatory compliance risk involving a material change. The changes to the gas pathway may potentially impact the safety of the device and quality of the gas condensates to the patient.
Disposable Patient Circuit that allows water to leak into the center gas lumen
There is a potential problem with the AXIOM Artis or Artis zee system in connection with the hand switch operated at the patient table which may result in uncontrolled release of an x-ray image.
Fresenius Naturalyte Liquid Bicarbonate Concentrate Product Number: 08-4000-LB The concentrate is formulated to be used with a three steam hemodialysis machine which is calibrated for acid and bicarbonate concentrates.
Fresenius Medical Care Holdings
Product was held at temperature above the labeled recommended storage temperature
Horiba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated. Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board stability claim and replacing it with a 1-day on-board stability claim.
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
Products were held at improper storage temperature
BD Vacutainer Citrate Tubes (LIGHT BLUE TOP TUBE 2.7ML)Part number: BD 363083 Part Number: FMC 87-3083-8 Single use tube used to collect, separate, transport or process blood specimens for laboratory determinations for in vitro diagnostic use. Coagulation studies.
Fresenius Medical Care Holdings
Products were held at improper storage temperature
Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.
Amerigel Hydrogel Saturated Gauze Dressing; 10 Individually Packaged pads size 2x2 4 Ply Pads; CAT # A5002. Wound management.
Amerx Health Care
Not approved labeling claims (antimicrobial and autolytic debridement)
TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.
Motion Concepts
Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).
Marketing without marketing clearance or approval.
The firm is conducting a field correction for the ADVIA Centaur and ADVIA Centaur XP Immunoassay systems, regarding manually loaded sample racks that are improperly loaded or pushed too far on the sample entry queue during normal operation of the ADVIA Centaur and ADVIA Centaur XP Immunoassay systems. Pushing the sample racks may cause misreads of Sample IDs (SID) if the sample rack loading instructions are not followed exactly as described in the ADVIA Centaur or ADVIA Centaur XP Operators Guide.