When a user switches on MR1100 Receiver, even though the battery was fully charged, the LED was displayed as orange color due to momentary drop of battery voltage.

Feb 15, 2017 Other Medical Devices Nationwide View Details →
Class I - Dangerous

The noncompliance is that the certification label was not affixed on the product, with statement "Complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007".

Jul 3, 2017 Surgical Instruments View Details →

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Feb 15, 2018 Other Medical Devices View Details →

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Feb 15, 2018 Other Medical Devices View Details →

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Mar 26, 2018 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

The noncompliance is that the certification label was not affixed on the product, with statement "Complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007".

Jul 3, 2017 Surgical Instruments View Details →

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Feb 15, 2018 Infusion Pumps View Details →

Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view.

Jan 26, 2018 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

The noncompliance is that the certification label was not affixed on the product, with statement "Complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007".

Jul 3, 2017 Surgical Instruments View Details →

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Mar 26, 2018 Diagnostic Equipment Nationwide View Details →

Potential errors in patient results generated by the Screening Center product that include both false negative and false positive results.

Nov 17, 2017 Diagnostic Equipment View Details →

Two lots of different sized modular heads potentially commingled. Risks include delay in surgery and dislocation due to use of an incorrect sized head.

Apr 2, 2018 Implants & Prosthetics Nationwide View Details →

A recent FDA inspection of the firm revealed significant Current Good Manufacturing Practices (cGMP) deficiencies including lack of data to support products conform to final specifications, and lack of expiration dates on the product being recalled.

Apr 25, 2018 Other Medical Devices View Details →

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Feb 15, 2018 Other Medical Devices View Details →