Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.
A change regarding the shelf life of a raw material was implemented without an appropriate submission of data demonstrating the continued efficacy of the product.
Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Module may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients.
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.