Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
SOMATOM Definition AS (Model 8098027) Computed tomography x-ray diagnostic system
Siemens Medical Solutions USA
There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM Definition Edge, SOMATOM Definition Flash and SOMATOM Force.
RICHARD WOLF TUBE SET FOR TEM, REF 4170.801. Used for the transanal endoscopic microsurgery.
Richard Wolf GmbH
Tube ends may be mixed up.
A subassembly component (dextrose bag) within this case product contained small amounts of sodium bicarbonate from a prior batch.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1004
Kelyniam Global
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
SOMATOM Definition Edge (Model 10590000) Computed tomography x-ray diagnostic system
Siemens Medical Solutions USA
There is a potential risk of unnecessary radiation exposure due to a software issue found in the CARE Dose4D algorithm implemented in Siemens Healthineers CT scanners of types SOMATOM Definition AS, SOMATOM Definition DS, SOMATOM Definition Edge, SOMATOM Definition Flash and SOMATOM Force.
A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.
There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.
5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 150mm shaft length, 2B5XT
Ethicon Endo-Surgery
Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.