SOMATOM Scope Power (Model 10967888)
Siemens Medical Solutions USA
A potential risk of unnecessary radiation exposure due to a software issue
Medical equipment, implants, and diagnostic devices
Siemens Medical Solutions USA
A potential risk of unnecessary radiation exposure due to a software issue
The drug dilution sequence for Moxifloxacin printed on the panel is incorrect. As a result, incorrect MIC results for Mxf could be reported.
GE Healthcare
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
ELITech Clinical Systems SAS
ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.
Underfilled reagent well that could lead to incorrect results
There is a potential for the knob assembly in the swivel adapter to fracture/break during use.
Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots.
The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.
Siemens Medical Solutions USA
A potential risk of unnecessary radiation exposure due to a software issue
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Zimmer Surgical
Devices may have a loose control bar, which could compromise the ability to control the thickness of the graft. Devices could also have a loose width plate that if undetected, could result in an imperfect, yet still usable graft. Potential risks include harvesting a skin graft of a thickness much greater than intended, described as patient gouged, minor delay of surgery, and sub-optimal (but usable) graft.
GE Healthcare
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Howmedica Osteonics
From March 2007 to October 2014, product did not contain updated instructions for use (IFU) to clarify the function of locking mechanism and warnings related to the locking screw.
Siemens Medical Solutions USA
A potential risk of unnecessary radiation exposure due to a software issue
A black, aluminum oxide coating may appear on the tray after cleaning. The black, aluminum oxide coating is not biocompatible.
GE Healthcare
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Product contains dry natural rubber latex. Label states Latex Free.
Siemens Medical Solutions USA
A potential risk of unnecessary radiation exposure due to a software issue
Baxter Healthcare
There were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.
The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.