🏥

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

38,779 Total Recalls

The drug dilution sequence for Moxifloxacin printed on the panel is incorrect. As a result, incorrect MIC results for Mxf could be reported.

May 4, 2018 Infusion Pumps View Details →
Class I - Dangerous

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

May 24, 2018 Other Medical Devices Nationwide View Details →

ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.

Apr 18, 2017 Other Medical Devices View Details →

There is a potential for the knob assembly in the swivel adapter to fracture/break during use.

Jun 29, 2018 Diagnostic Equipment Nationwide View Details →

Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots.

Mar 1, 2018 Diagnostic Equipment Nationwide View Details →

The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.

May 16, 2018 Diagnostic Equipment View Details →
Class I - Dangerous

Devices may have a loose control bar, which could compromise the ability to control the thickness of the graft. Devices could also have a loose width plate that if undetected, could result in an imperfect, yet still usable graft. Potential risks include harvesting a skin graft of a thickness much greater than intended, described as patient gouged, minor delay of surgery, and sub-optimal (but usable) graft.

Jun 8, 2018 Other Medical Devices View Details →

A black, aluminum oxide coating may appear on the tray after cleaning. The black, aluminum oxide coating is not biocompatible.

Jun 26, 2018 Surgical Instruments View Details →
Class I - Dangerous

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

May 24, 2018 Other Medical Devices Nationwide View Details →

The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.

Jun 8, 2018 Diagnostic Equipment View Details →