In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.

Jun 29, 2018 Diagnostic Equipment Nationwide View Details β†’

GE Healthcare Optima CT660. Sold under the following product names: BEIJING OPTIMA CT660 PRO GT170, BEIJING OPTIMA CT660 PRO GT200, BEIJING OPTIMA CT660S PRO 32 S, CT GOLDSEAL, CT GOLDSEAL OPTIMA 660, CT GOLDSEAL OPTIMA 660 UT-BUC, CT OPTIMA 660 LNG HN P20, CT OPTIMA 660 MID HN P20, CT OPTIMA 660 PRO, CT OPTIMA 660 PRO LONG BJG P2., CT OPTIMA 660 PRO MID, CT OPTIMA 660 PRO MID HN P20, CT OPTIMA 660S LNG HN P20, CT OPTIMA CT660 GT 1700, CT OPTIMA CT660 GT1700, CT OPTIMA CT660 GT1700V, CT OPTIMA CT660 PRO GT1700, CT OPTIMA CT660 PRO LNG P1.5 H, CT OPTIMA CT660 PRO MID P1.5 H, CT OPTIMA CT660 SE GT 1700, CT OPTIMA CT660 SE GT2000, CT660, CT660 ASIR OPT, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT2000 MKE25, OPTI660_A_G_1.5, OPTI660_A_L_2.0, OPTI660_N_G_1.5, OPTI660_N_G_2.0, OPTI660_N_L_2.0, OPTIMA 660 DIVA, OPTIMA 660 LNG HN P20, OPTIMA 660 LNG HN P20, OPTIMA 660 LONG, OPTIMA 660 LONG BJG P2.0, OPTIMA 660 LONG WSO, OPTIMA 660 MID, OPTIMA 660 MID BJG P1.5, OPTIMA 660 MID HN P20, OPTIMA 660 MID HN P20, OPTIMA 660 MID WSO, OPTIMA 660 PRO LGN HN P20, OPTIMA 660 PRO LONG BJG P1.5_, OPTIMA 660 PRO LONG BJG P2, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG WSO, OPTIMA 660 PRO MID BJG P1.5, OPTIMA 660 PRO MID BJG P2.0, OPTIMA 660 PRO MID H, OPTIMA 660 PRO MID HN P20, OPTIMA 660 PRO MID WSO, OPTIMA 660 WSO, OPTIMA 660S 32 OPTIMA 660S 32 SLICE LONG WSO, OPTIMA 660S 32 SLICE MID WSO, OPTIMA 660S LNG HN P20, OPTIMA 660S MID BJG P2.0, OPTIMA 660S PRO 32 SLICE M WSO, OPTIMA 660S PRO LONG BJG P2.0, OPTIMA 660S PRO MID BJG P1.5, OPTIMA 660S WSO, OPTIMA CT 660 FREEDOM, OPTIMA CT660, OPTIMA CT660 2ND EDITION, OPTIMA CT660 ASIR, OPTIMA CT660 ASIR ADV 2ND, OPTIMA CT660 FREEDOM, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG HN2, OPTIMA CT660 FREEDOM LG HN2.5, OPTIMA CT660 FREEDOM MD, OPTIMA CT660 FREEDOM MD BJG2., OPTIMA CT660 GS, OPTIMA CT660 GT 1700 BJG2, OPTIMA CT660 GT 1700 BJG2.5, OPTIMA CT660 GT 1700 HN P2, OPTIMA CT660 GT 1700 HN P2., OPTIMA CT660 GT 1700 HN P2.5, OPTIMA CT660 GT 2000 HN P2.5, OPTIMA CT660 GT1700, OPTIMA CT660 GT1700 MKE2.5, OPTIMA CT660 GT1700 PHASE 1.5, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000 PHASE 1.5, OPTIMA CT660 LITE, OPTIMA CT660 LITE HN P2.5, OPTIMA CT660 MID P1.5 HINO, OPTIMA CT660 MID P20, OPTIMA CT660 MID P2000, OPTIMA CT660 PRO, OPTIMA CT660 PRO GT 2000, OPTIMA CT660 PRO GT1700, OPTIMA CT660 PRO GT1700 ASIR, OPTIMA CT660 PRO GT1700 P20, OPTIMA CT660 PRO GT1700 PHASE, OPTIMA CT660 PRO GT1700 PHASE 1.5, OPTIMA CT660 PRO GT1700V, OPTIMA CT660 PRO GT1700V ASIR, OPTIMA CT660 PRO GT2000, OPTIMA CT660 PRO GT2000 PHASE, OPTIMA CT660 PRO GT2000 PHASE 1.5, OPTIMA CT660 PRO LNG, OPTIMA CT660 PRO LNG P1.5, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LONG BJG P2.0, OPTIMA CT660 PRO LONG P2.0, OPTIMA CT660 PRO MID P1.5, OPTIMA CT660 PRO MID P1.5 H, OPTIMA CT660 PRO MID P1.5 HINO, OPTIMA CT660 SE, OPTIMA CT660 SE GT 1700 BJG, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 HN2, OPTIMA CT660 SE GT 2000, OPTIMA CT660 SE GT 2000 BJG, OPTIMA CT660 SE GT 2000 BJG2., OPTIMA CT660 SE GT 2000 BJG2.5, OPTIMA CT660 SE GT 2000 HN2, OPTIMA CT660 SE GT 2000 HN2.5, OPTIMA CT660 SE GT2000 MKE2.5, OPTIMA CT660 SE WITH ASIR, OPTIMA CT660 W/PRO PKG-BR, OPTIMA CT660 W/PRO PKG-LA, OPTIMA CT660-64 USA, OPTIMA CT660S, OPTIMA CT660S, OPTIMA CT660S 32 SLICE GT1700, OPTIMA CT660S 32 SLICE HINO, OPTIMA CT660S GT, OPTIMA CT660S GT 1700 HN2.5, OPTIMA CT660S GT 2000, OPTIMA CT660S GT 2000 BJG2.5, OPTIMA CT660S GT 2000 HN2.5, OPTIMA CT660S GT1700 BJG P2.5, OPTIMA CT660S GT1700 HN P2.5, OPTIMA CT660S PRO 32 HINO, OPTIMA CT660S PRO 32 SLICE, OPTIMA CT660S PRO 32 SLICE GT2, OPTIMA CT660S PRO P1.5 HINO, OPTIMACT660_A_G, OPTIMACT660_A_L, OPTIMACT660_N_G, OPTIMACT660_N_L, REVOLUTION EVO, REVOLUTION EVO WITH GT2000 TABLE (H

GE Healthcare Japan

Class I - Dangerous

GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.

Apr 27, 2018 Infusion Pumps Nationwide View Details β†’

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.

Jul 13, 2018 Diagnostic Equipment Nationwide View Details β†’

Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.

Feb 27, 2017 Diagnostic Equipment Nationwide View Details β†’

CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, right plates were laser marked as left plates and the left plates were laser marked as right plates.

Jul 16, 2018 Implants & Prosthetics Nationwide View Details β†’

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.

Jul 13, 2018 Diagnostic Equipment Nationwide View Details β†’

CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, right plates were laser marked as left plates and the left plates were laser marked as right plates.

Jul 16, 2018 Implants & Prosthetics Nationwide View Details β†’