The terminals on the plug may become loose and overheat.

Sep 28, 2018 Diagnostic Equipment Nationwide View Details →

The conduit length and branch length of a SG Pulmonary Hemi-Artery were switched on the product labeling. The graft labeling indicates the conduit length is 1.0cm and the branch length is 4.5cm.

Sep 6, 2018 Other Medical Devices Nationwide View Details →

These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip

Aug 21, 2018 Infusion Pumps Nationwide View Details →

The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.

Aug 30, 2018 Diagnostic Equipment View Details →