Patient monitors may restart due to network overload caused by network configuration.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
CIVCO Needle Guide Bracket, REF 674-057, Reusable non-sterile bracket for use with SonoSite L25 series tranducers, Rx.
CIVCO Medical Instruments Co.
There is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.
A percentage of the products shipped in Smartscope single-box packaging may exhibit a broken package seal, thus having compromised sterility. There have been two reports of compromised sterility; however, all products packaged in Smartscope single-box packaging are at risk of a compromised product seal. Broken seal pouches compromise sterility and could increase the risk of infection if used.
SonoSite L25 Bracket Kit containing 1 Reusable non-sterile bracket, REF numbers P03122-01, P03122-02, P03122-03, P03122-04, and P03122-05, Mfg for: SonoSite, Inc., Bothell, WA.
CIVCO Medical Instruments Co.
The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
Patient monitors may restart due to network overload caused by network configuration.
Due to some cards in this lot experiencing variations in wave form, resulting in some cards with shorter or longer times than expected
Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
Discovery NM/CT 670 Pro, model 5376204-70-54
GE Healthcare
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
BioFlo PICC with ENDEXO and PASV, ( RS 5F DL BIOFLO PASV - MEMORIAL HOSPITAL SYSTEM (LTP) PG) Cat. no. 60M183481
Angiodynamics Inc. (Navilyst Medical Inc.)
specific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by validation testing. Although the product under recall has not yet reached its validated expiry date, removal of the affected product is warranted to prevent use beyond the validated expiry date.
Patient monitors may restart due to network overload caused by network configuration.
CIVCO Needle Guide Starter Kit, REF 674-038, Reusable non-sterile bracket with sterile needle guide (1cm, 1.5cm, 2cm depths) with (10.2 x 147cm) CIV-Flex cover for use with SonoSite L25 series transducers, 21 GAUGE, Rx. The firm name on the label is CIVCO, Kalona, IA..
CIVCO Medical Instruments Co.
Comprehensive Mini Stem, Item No. 113631:
Zimmer Biomet
Zimmer Biomet is conducting a lot specific medical device recall for two lots of the Orthopedic Salvage System (OSS) and the Comprehensive Shoulder System. The investigation determined that the Mini Humeral Stem is potentially labeled as the Bowed IM Stem lot. The associated risk of this product issue is a potential extension of surgery <30 minutes if a replacement is readily available.
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
CIVCO Needle Guide Bracket, REF P03122-06, Reusable non-sterile bracket for use with SonoSite L25/L25e transducers, Mfg by: CIVCO, Kalona, IA.
CIVCO Medical Instruments Co.
The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available.
CIVCO Needle Guide Bracket, REF numbers M001457100, P06013-10, and P03122-08, reusable non-sterile single-angle bracket for use with SonoSite L25 series tranducers, RX. The firm name on the label is CIVCO, Kalona, IA.
CIVCO Medical Instruments Co.
The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.