Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Feb 25, 2026 Surgical Instruments Nationwide View Details →

16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.

Feb 13, 2026 Surgical Instruments Nationwide View Details →

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Feb 25, 2026 Surgical Instruments Nationwide View Details →

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Feb 25, 2026 Surgical Instruments Nationwide View Details →

16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.

Feb 13, 2026 Surgical Instruments Nationwide View Details →

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Feb 25, 2026 Surgical Instruments Nationwide View Details →

16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.

Feb 13, 2026 Surgical Instruments Nationwide View Details →

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Jan 8, 2026 Surgical Instruments Nationwide View Details →

Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081

Medline Industries

Class I - Dangerous

Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.

Dec 31, 2025 Surgical Instruments Nationwide View Details →

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Jan 7, 2026 Surgical Instruments Nationwide View Details →

Medline Surgical Gowns: 1) GOWN,NONRNF,L,30/CS, Model Number: DYNJP2001; 2) GOWN,SIRUS,NONRNF,SETINSLV,L,20/CS, Model Number: DYNJP2001S; 3) GOWN,ECLIPSE,NONRNF,XL,ST,30/CS, Model Number: DYNJP2002; 4) GOWN,SIRUS,NONRNF,SETINSLV,XL,20/CS, Model Number: DYNJP2002S; 5) GOWN,SIRUS,NONRNF,XLN/XL,20/CS, Model Number: DYNJP2002SL; 6) GOWN,NONRNF,2XL,18/CS, Model Number: DYNJP2003; 7) GOWN,SIRUS,NONRNF,SETINSLV,2XL,18/CS, Model Number: DYNJP2003S; 8) GOWN,SIRUS,NONRNF,XLN/2XL,18/CS, Model Number: DYNJP2003SL; 9) GOWN,NONRNF,3XL,18/CS, Model Number: DYNJP2004; 10) GOWN,SIRUS,NONRNF,3XL,18/CS, Model Number: DYNJP2004S; 11) GOWN,NONRNF,SM,30/CS, Model Number: DYNJP2005; 12) GOWN,SIRUS,NONRNF,SM,30/CS, Model Number: DYNJP2005S; 13) GOWN,NONRNF,4XL,18/CS, Model Number: DYNJP2009; 14) GOWN,SIRUS,NONRNF,4XL,18/CS, Model Number: DYNJP2009S; 15) GOWN,ECLIPSE,FABRNF,L,30/CS, Model Number: DYNJP2101; 16) GOWN,SIRUS,FABRNF,L,20/CS, Model Number: DYNJP2101S; 17) GOWN,ECLIPSE,FABRNF,XL,30/CS, Model Number: DYNJP2102; 18) GOWN,SIRUS,FABRNF,XL,20/CS, Model Number: DYNJP2102S; 19) GOWN,ECLIPSE,FABRNF,2XL,18/CS, Model Number: DYNJP2103; 20) GOWN,SIRUS,FABRNF,2XL,18/CS, Model Number: DYNJP2103S; 21) GOWN,ECLIPSE,POLYRNF,L,30/CS, Model Number: DYNJP2201; 22) GOWN,SIRUS,POLYRNF,SETINSLV,L,20/CS, Model Number: DYNJP2201S; 23) GOWN,ECLIPSE,POLYRNF,XL,30/CS, Model Number: DYNJP2202; 24) GOWN,SIRUS,POLYRNF,SETINSLV,XL,20/CS, Model Number: DYNJP2202S; 25) GOWN,ECLIPSE,POLYRNF,2XL,18/CS, Model Number: DYNJP2203; 26) GOWN,SIRUS,POLYRNF,SETINSLV,2XL,18/CS, Model Number: DYNJP2203S; 27) GOWN,ECLIPSE,POLYRNF,XLN/L,28/CS, Model Number: DYNJP2204; 28) GOWN,SIRUS,POLYRNF,XLN/LG,20/CS, Model Number: DYNJP2204S; 29) GOWN,ECLIPSE,POLYRNF,XLN/XL,28/CS, Model Number: DYNJP2205; 30) GOWN,SIRUS,POLYRNF,XLN/XL,20/CS, Model Number: DYNJP2205S; 31) GOWN,ECLIPSE,POLYRNF,XLN/2XL,18/CS, Model Number: DYNJP2206; 32) GOWN,SIRUS,POLYRNF,XLN/2XL,18/CS, Model Number: DYNJP2206S; 33) DBQ-GOWN,PREVENTION PLUS,LN/L,ST,24/CS, Model Number: DYNJP2301P; 34) DBQ-GOWN,PREVENTION PLUS,XLN/XL,ST,24/CS, Model Number: DYNJP2302P; 35) DBD-GOWN,PREVENTION PLUS,XLN/2XL,ST,22/C, Model Number: DYNJP2303P; 36) DBD-GOWN,PREVENTION PLUS,L,ST,24/CS, Model Number: DYNJP2306P; 37) DBQ-GOWN,PREVENTION PLUS,XL,ST,24/CS, Model Number: DYNJP2307P; 38) DBQ-GOWN,PREVENTION PLUS,2XL,ST,22/CS, Model Number: DYNJP2308P; 39) GOWN,ORBIS,LVL 3,LRG/XLONG,ST,24/CS, Model Number: DYNJP2361P; 40) GOWN,ORBIS,LVL 3,XLONG/XLARGE,ST,24/CS, Model Number: DYNJP2362P; 41) GOWN,ORBIS,LVL 3,XLNG/XXLARGE,ST,22/CS, Model Number: DYNJP2363P; 42) GOWN,ORBIS,LVL 3,LARGE,ST,24/CS, Model Number: DYNJP2366P; 43) GOWN,ORBIS,LVL 3,XLARGE,ST,24/CS, Model Number: DYNJP2367P; 44) GOWN,ORBIS,LVL 3,XXLARGE,ST,22/CS, Model Number: DYNJP2368P; 45) GOWN,ORBIS,LVL 3,4XL,ST,20/CS, Model Number: DYNJP2369P; 46) GOWN,SIRUS,NONRNF,RAGLAN,L,ST,32/CS, Model Number: DYNJP2401; 47) GOWN,SIRUS,NONRNF,RAGLAN,XL,ST,30/CS, Model Number: DYNJP2402; 48) GOWN,SIRUS,NONRNF,RAGLAN,2XL,ST,28/CS, Model Number: DYNJP2403; 49) GOWN,SIRUS,FABRNF,RAGLAN,L,ST,30/CS, Model Number: DYNJP2501; 50) GOWN,SIRUS,FABRNF,RAGLAN,XL,ST,28/CS, Model Number: DYNJP2502; 51) GOWN,SIRUS,FABRNF,RAGLAN,2XL,ST,28/CS, Model Number: DYNJP2503; 52

Medline Industries

Class I - Dangerous

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Jan 7, 2026 Surgical Instruments Nationwide View Details →

Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) L & D PACK-LF, Model Number: DYNJ0456816O; 6) L & D PACK-LF, Model Number: DYNJ0456816P; 7) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 8) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 10) DELIVERY PACK-LF, Model Number: DYNJ20546F; 11) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 13) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 14) OB PACK-LF, Model Number: DYNJ26121I; 15) OB DELIVERY PACK, Model Number: DYNJ31024J; 16) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 17) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 18) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 19) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 20) VAG DELIVERY PACK, Model Number: DYNJ38481G; 21) VAG DELIVERY PACK, Model Number: DYNJ38481I; 22) VAG DELIVERY PACK, Model Number: DYNJ38481K; 23) OB PACK 2, Model Number: DYNJ39000D; 24) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 25) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 26) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 27) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 28) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 29) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 31) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 32) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 33) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 34) OB-VAG PACK 9128190, Model Number: DYNJ62341; 35) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 36) L & D PACK, Model Number: DYNJ63216A; 37) L&D PACK, Model Number: DYNJ63757A; 38) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 39) MINOR VAG PACK, Model Number: DYNJ68663; 40) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 41) GYN VAGINAL PACK, Model Number: DYNJ68681; 42) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 43) VAGINAL PACK, Model Number: DYNJ69774A; 44) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 45) DELIVERY PACK, Model Number: DYNJ87469A; 46) VAGINAL DELIVERY, Model Number: DYNJ901387A; 47) FLOYD DELIVERY, Model Number: DYNJ906937D; 48) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 49) OB PACK, Model Number: DYNJT5186; 50) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 51) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 52) LABOR AND DELIVERY KIT, Model Number: MNS8950A

Medline Industries

Class I - Dangerous

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Jan 7, 2026 Surgical Instruments Nationwide View Details →