Surgical Instruments

🏥 Medical Devices 6,839 recalls

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

Dec 15, 2015 Surgical Instruments Nationwide View Details →

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

Dec 15, 2015 Surgical Instruments Nationwide View Details →

Stryker Sustainability Solutions is recalling the Fuhrman Pleural & Pneumopericardial Drainage Set due to receiving complaints that it broke off in the pleural cavity while inserting the device into the patient.

Nov 17, 2015 Surgical Instruments View Details →

The firm discovered that information printed on the quick reference magnet is incorrect; specifically, Fahrenheit to Celsius temperature conversions and the default Warming Cabinet temperature.

Nov 20, 2015 Surgical Instruments Nationwide View Details →

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSCs product risk documentation.

Nov 20, 2015 Surgical Instruments Nationwide View Details →

Out of tolerance for radio frequency emissions. At the 150-1000MHz frequency, the testing indicated the BrightMatter Navigation system was up to 20dB uV/meter higher than the applicable IEC 60601-1-2:2007 (Ed3.0) standard specification.

Dec 4, 2015 Surgical Instruments View Details →

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSC's product risk documentation.

Nov 20, 2015 Surgical Instruments Nationwide View Details →

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSC's product risk documentation.

Nov 20, 2015 Surgical Instruments Nationwide View Details →

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSCs product risk documentation.

Nov 20, 2015 Surgical Instruments Nationwide View Details →

Two lots of oscillating saw blades incorrectly identify the blade cutting width as 25mm instead of the correct specification of 19mm.

Dec 8, 2015 Surgical Instruments View Details →