Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected.

Feb 28, 2014 Other Medical Devices Nationwide View Details →

Artis Zee Angiographic x-ray systems.

Siemens Medical Solutions USA

Class I - Dangerous

An inappropriate cable connection was assembled for certain Artis zee systems. which may cause a system failure and possibly lead to a restricted maneuverability of the gantry system.

Feb 7, 2014 Other Medical Devices View Details →

The Lock Nut component of the Jet appliances may be missing the screw collar, which serves as a mechanical stop to prevent over-tightening in appliance assemby. Without this collar, an over-tightened lock nut may stop the piston assembly from moving freely, subsequently preventing intended movement of the teeth.

Feb 24, 2014 Other Medical Devices Nationwide View Details →

Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may lead an orthodontist to place the bracket on the wrong tooth, leading to movement 10 degrees in the wrong direction.

Mar 25, 2014 Other Medical Devices View Details →