URO ROBOTIC PACK , Model No LVUR44C LVUR44C-03
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
π₯ Medical Devices β’ 15,203 recalls
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
Waldemar Link GmbH & Co. KG (Mfg Site)
A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
Under certain circumstances, it can occur that patient records from unrelated individuals are automatically merged into one patient record without user notification.
Smiths Medical ASD
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
Beckman Coulter
Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
Beckman Coulter
Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
FUJIFILM Healthcare Americas
The incorrect computed patient age is showing in VX for patients less than 3 months old.
MEDLINE INDUSTRIES, LP - Northfield
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
MEDLINE INDUSTRIES, LP - Northfield
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Angiodynamics
Product is mislabeled with the incorrect fill volume.
Waldemar Link GmbH & Co. KG (Mfg Site)
A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
American Contract Systems
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.