Other Medical Devices

🏥 Medical Devices • 15,809 recalls

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

Syngo.Via VB20A model 1049610

Siemens Medical Solutions USA

Class I - Dangerous

Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.

Mar 27, 2018 Other Medical Devices Nationwide View Details →

The firm was marketing the Ondamed System in the US without marketing clearance. As a corrective action, the firm revised its device design to comply with regulations. The firm has notified customers that an upgrade for the device is now available, which includes the addition of direct readout and audio signals for heart rate, and visual patient readout of blood oxygen content SpO2.

Jul 7, 2017 Other Medical Devices Nationwide View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

ADVIA Centaur¿ HBc IgM

Siemens Healthcare Diagnostics

Class I - Dangerous

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Jan 2, 2018 Other Medical Devices Nationwide View Details →

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Oct 19, 2017 Other Medical Devices View Details →

IMMULITE ¿ /IMMULITE ¿ 1000 CEA

Siemens Healthcare Diagnostics

Class I - Dangerous

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Jan 2, 2018 Other Medical Devices Nationwide View Details →

IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi CEA

Siemens Healthcare Diagnostics

Class I - Dangerous

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Jan 2, 2018 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Jan 2, 2018 Other Medical Devices Nationwide View Details →

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.

Jan 2, 2018 Other Medical Devices Nationwide View Details →