Intera CV, Model Number: 781107;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class I - Dangerous

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

Mar 24, 2025 Other Medical Devices Nationwide View Details →

The perc pin adaptor AMCH07200 tightening knob was found to be inconvenient to the user and the tightening markings found to be improperly placed.

Apr 14, 2025 Other Medical Devices Nationwide View Details →

Disconnection of the adapter from the endotracheal tube can potentially be difficult or not possible.

Apr 7, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.

Apr 8, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Apr 2, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Apr 2, 2025 Other Medical Devices Nationwide View Details →

In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

Mar 27, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Mar 21, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Mar 21, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Wrong analysis results can occur if the user creates two structures with the same name, except for one differing by a special character.

Apr 1, 2025 Other Medical Devices Nationwide View Details →

The device was delivered with a preassembled Femoral Augment screw that was missing its thread.

Apr 3, 2025 Other Medical Devices Nationwide View Details →