Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Other Medical Devices
🏥 Medical Devices • 15,203 recalls
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Cardiac Pack - LF, Model DYNJ31903I - Product Usage: This is a type of convenience kit used in various surgeries including cardiac, abdominal, and neck surgeries.
Medline Industries
The cylindrical sponge component is not x-ray detectable.
The cylindrical sponge component is not x-ray detectable.
Thyroid Pack CKHA-LF, Model DYNJ50352B - Product Usage: This is a type of convenience kit used in various surgeries including cardiac, abdominal, and neck surgeries.
Medline Industries
The cylindrical sponge component is not x-ray detectable.
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Manufacturing packaging error.
Simple Neck Dissection, Model DYNJ47481C - Product Usage: This is a type of convenience kit used in various surgeries including cardiac, abdominal, and neck surgeries.
Medline Industries
The cylindrical sponge component is not x-ray detectable.
Purple ink located on the outside label of Vis-U-All Pouches may migrate (bleed) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
The device was assembled with the incorrect exhalation port which can cause gas leakage.
cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clinical use
Roche Diagnostics Operations
During the subsequent pipetting (1st or 2nd measurement), the residual reagent can sporadically be transferred from the conus into the reaction cuvette and lead to inaccurately low results for the affected TDMs (Therapeutic Drug Monitoring).
Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for clinical use Part Number: 05641446001
Roche Diagnostics Operations
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
Roche cobas 8000 modular analyzer series (i.e., cobas 8000 ISE module; cobas c 502, 701, and 702 modules; and cobas e 602 and 801 modules). Chemistry Analyzer (photometric discrete), for clinical use Part Number: 08763662001
Roche Diagnostics Operations
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.