🏨

Other Medical Devices

πŸ₯ Medical Devices β€’ 15,203 recalls

The text on the patient pack is correct and the product itself remains unchanged. FlexMedics does not expect device failures; Instead, there is an opportunity for misuse of the device if the label is not read. If the band is too small, the treatment may be more aggressive. It is important to note that elastic band force decreases over time. If the band is too small and still applied, the band could break. Both circumstances could result in temporary patient discomfort.

Jan 21, 2021 Other Medical Devices Nationwide View Details β†’

Non-sterile product was shipped to customers labelled as sterile.

Jan 19, 2021 Other Medical Devices Nationwide View Details β†’

The affected Milex Gellhorn pessaries were incorrectly manufactured with a long stem (2.170 ), instead of a short stem (1.670 or 1.960 nominal height) as required. The length discrepancy does not meet the released print dimensions.

Jan 21, 2021 Other Medical Devices Nationwide View Details β†’

The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins.

Jan 12, 2021 Other Medical Devices Nationwide View Details β†’

Slits were identified in a low quantity of ORS-320 Slush + Warmer Disc Drapes from a single lot. A source for potential manufacturing error was identified and investigation is underway to prevent further recurrence. The slits may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment. This could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.

Jan 11, 2021 Other Medical Devices Nationwide View Details β†’

incorrect device is contained in the package. The packaging label indicates that the device is for left side use, the product inside the package is for right side use. Use of an opposite side mesh can result in implant failure and hernia recurrence

Jan 13, 2021 Other Medical Devices Nationwide View Details β†’

A deformity affecting a specific lot of VITROSΒΏ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.

Jan 14, 2021 Other Medical Devices Nationwide View Details β†’

GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.

Dec 17, 2020 Other Medical Devices Nationwide View Details β†’

IOLMaster 700

Carl Zeiss Meditec AG

Class I - Dangerous

When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen

Dec 7, 2020 Other Medical Devices Nationwide View Details β†’

Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus

Jan 11, 2021 Other Medical Devices Nationwide View Details β†’

Medtronic PS Medical Cerebro Spinal Fluid (CSF)-Flow Control Valves for the following CFNs: 22011 L- SHUNT 22011 L INTL ONLY ULTRA SM LOW 22011 LL - SHUNT 22011 LL INTL ONLY ULTRA SM L-L 22011 M - SHUNT 22011 M INTL ONLY ULTRA SM MED 22017 B-L - SHUNT 22017 B-L ASSY US PER 90CM SM LOW 22017 B-M - SHUNT 22017 B-M ASSY US PER 90CM SM MED 22017B-LL - SHUNT 22017B-LL ASSY US PER 90CM SM LL 23093 - VALVE 23093 CONT. REG. LOW-LOW 27102 - VALVE 27102 CONTOURED SML LOW LOW 27553 M - VALVE 27553 M CSF-FLOW CNTRL SM EXTR MP 27643M - SHUNT 27643M ASSY SM PCATH130CM MED 27683M - SHUNT 27683M RG SNP PCATH90 PLI STD OE 42312 - VALVE 42312 SM CONTOURED LOW PRES IMP 42314- VALVE 42314 SM CONTOURED MED PRESS IMP 42316 - VALVE 42316 SM CONTOURED HIGH PRESS 42322 - VALVE 42322 CONTOURED REG LOW PRESS IMP 42324 - VALVE 42324 CONTOURED REG MED PRESS IMP 42326 - VALVE 42326 CONT HIGH PRESS IMP 42410 - VALVE 42410 ULTRA SMALL LOW-LOW 42414 - VALVE 42414 ULTRA SMALL MEDIUM PRESS IMP 42532 - VALVE 42532 BURR HOLE 12MM LOW PRESS IMP 42534 - VALVE 42534 BURR HOLE 12MM MED PRESS IMP 42536 - VALVE 42536 BURR HOLE 12MM HI PRESS IMP 42542 - VALVE 42542 BURR HOLE 16 MM LOW IMP 42544 - VALVE 42544 BURR HOLE 16MM MED PRESS IMP 42546 - VALVE 42546 BURR HOLE 16MM HI PRESS IMP 46564 - SHUNT 46564 CSF SNAP ASSY US MEDIUM 46622 - SHUNT 46622 CSF ASSY SMALL LOW CONT C/P 46624 - SHUNT 46624 CSF ASSY SMALL MED CONT C/P 46626 - SHUNT 46626 CSF ASSY SMALL HIGH CONT C/P 46642 - SHUNT 46642 ASSY REG LOW C/P-90CM 46644 - SHUNT 46644 ASSY REG MED C/P-90CM 46646 - SHUNT 46646 ASSY REG HIGH C/P-90CM 9003 A - KIT 9003 A INTL ONLY SHUNT CONT SM LO 9003 B - KIT 9003 B INTL ONLY SHUNT CONT SM ME 9003 C - KIT 9003 C INTL ONLY SHUNT CONT SM HI 9003 D - KIT 9003 D SHUNT INTL ONLY CONT REG L 9003 E - KIT 9003 E SHUNT INTL ONLY CONT REG M 9003 F - KIT 9003 F SHUNT INTL ONLY CONT REG H 9040 A - KIT 9040 A SHUNT INTL ONLY BH 12MM LOW 9040 B - KIT 9040 B SHUNT INTL ONLY BH 12MM MED 9040 C - KIT 9040 C SHUNT INTL ONLY BH 12MM HIGH 9040 D - KIT 9040 DSHUNT INTL ONLY BH 16MM LOW 9040 E - KIT 9040 E SHUNT INTL ONLY BH 16MM MED 9040 F - KIT 9040 A SHUNT INTL ONLY BH 12MM LOW 92322 - VALVE 92322 CSF CONT REG LOW BIO 92324 - VALVE 92324 CSF CONT REG MED BIO

Medtronic Neurosurgery

Class I - Dangerous

Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus

Jan 11, 2021 Other Medical Devices Nationwide View Details β†’

Medtronic Delta CSF (Cerebro Spinal Fluid) -Flow Control Valves for the following CFNs/REF numbers: 24048 - VALVE 24048 DELTA RGL PL 0.5, 24053 - CHAMBER 24053 DELTA, 24147 - VALVE 24147 DELTA SML PL0.5, 25121-1 - ASSY 25121-1 DELTA SML SNAP PERIT 120, 25121-5 - ASSY 25121-5 DELTA SML SNAP P-CATH 120, 25131-1 - DELTA 25131-1 INTL ONLY SHUNT SM 1.0, 25131-2 - SHUNT 25131-2 INTL ONLY DELTA SM 2.0, 25131-5 - SHUNT 25131-5 INTL ONLY DELTA SM 1.5, 25132-2 - SHUNT 25132-2 INTL ONLY DELTA REG.2.0, 25132-5 - SHUNT 25132-5 INTL ONLY DELTA REG 1.5, 27219-1 - VALVE 27219-1 NEONATE DELTA, 27219-2 - VALVE 27219-2 NEONATE DELTA, 27219-5 - VALVE 27219-5 DELTA NEONATE 1.5 IMP, 27554-1 - VALVE 27554-1 DELTA SM EXTRACTED PL 1.0, 27554-2 - VALVE 27554-2 DELTA SM EXTRACTED PL 2.0, 27554-5 - VALVE 27554-5 DELTA SM EXTRACTED PL 1.5, 42812 - VALVE 42812 DELTA SMALL LEVEL I, 42813 - VALVE 42813 DELTA SMALL LEVEL 1.5, 42814 - VALVE 42814 DELTA SMALL LEVEL II, 42822 - VALVE 42822 DELTA REG PRESS LEVEL I IMP, 42823 - VALVE 92823 BIOGLIDE DELTA REG. 1.5, 42824 - VALVE 42824 DELTA REG PRESS LEVEL II IMP, 46812 - ASSY 46812 DELTA SHUNT SMALL PL 1, 46813 - ASSY 46813 DELTA SHUNT SM LEV 1.5, 46822 - ASSY 46822 DELTA SHUNT REG P/L 1, 46823 - ASSY 46823 DELTA SHUNT REG LEV 1.5, 46824 - ASSY 46824 DELTA SHUNT REG P/L 2, 46832 - SHUNT 46832 NEONATAL DELTA SNAP L 1, 46833 - SHUNT 46833 NEONATAL DELTA SNAP 1.5, 46834 - SHUNT 46834 NEONATAL DELTA SNAP L 2, 92822 - VALVE 92822 BIOGLIDE DELTA REG. 1.0, 92823 - VALVE 92823 BIOGLIDE DELTA REG. 1.5,

Medtronic Neurosurgery

Class I - Dangerous

Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus

Jan 11, 2021 Other Medical Devices Nationwide View Details β†’

There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.

Jan 7, 2021 Other Medical Devices Nationwide View Details β†’