TECNIS Toric 1-Piece IOL, Model Number: ZCT150 +23.5D SE 1.50D CYL - Product Usage: intended to be placed in the capsular bag.
Johnson & Johnson Surgical Vision
Due to the release of nonconforming Intraocular Lenses (IOLs).
π₯ Medical Devices β’ 15,203 recalls
Johnson & Johnson Surgical Vision
Due to the release of nonconforming Intraocular Lenses (IOLs).
The text on the patient pack is correct and the product itself remains unchanged. FlexMedics does not expect device failures; Instead, there is an opportunity for misuse of the device if the label is not read. If the band is too small, the treatment may be more aggressive. It is important to note that elastic band force decreases over time. If the band is too small and still applied, the band could break. Both circumstances could result in temporary patient discomfort.
Non-sterile product was shipped to customers labelled as sterile.
The affected Milex Gellhorn pessaries were incorrectly manufactured with a long stem (2.170 ), instead of a short stem (1.670 or 1.960 nominal height) as required. The length discrepancy does not meet the released print dimensions.
The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins.
Slits were identified in a low quantity of ORS-320 Slush + Warmer Disc Drapes from a single lot. A source for potential manufacturing error was identified and investigation is underway to prevent further recurrence. The slits may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment. This could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.
incorrect device is contained in the package. The packaging label indicates that the device is for left side use, the product inside the package is for right side use. Use of an opposite side mesh can result in implant failure and hernia recurrence
A deformity affecting a specific lot of VITROSΒΏ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.
There is the potential of a bent or disconnected needle in one of the components of the convenience kits.
GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.
Carl Zeiss Meditec AG
When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen
Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus
Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus
Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus
GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.
Combat Medical Systems
There is the potential of a bent or disconnected needle in one of the components of the convenience kits.
Combat Medical Systems
There is the potential of a bent or disconnected needle in one of the components of the convenience kits.
Aomori Olympus Co.
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
Aomori Olympus Co.
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.