🏨

Other Medical Devices

🏥 Medical Devices • 15,809 recalls

There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.

Feb 4, 2022 Other Medical Devices Nationwide View Details →
Class I - Dangerous

The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.

Jun 30, 2021 Other Medical Devices Nationwide View Details →
Class I - Dangerous

It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Jan 10, 2022 Other Medical Devices Nationwide View Details →

YSIO X.Pree

Siemens Medical Solutions USA

Class I - Dangerous

For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.

Mar 7, 2022 Other Medical Devices Nationwide View Details →

Hanger-Bar 2P 45cm

Human Care USA

Class I - Dangerous

The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.

Nov 1, 2021 Other Medical Devices Nationwide View Details →