The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.
Other Medical Devices
đĽ Medical Devices ⢠15,203 recalls
SJM Pericardial Patch, with EnCap Technology 9x14cm, Model Number REF C0914; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE
Glycar SA Pty.
The impacted lot may not meet the required tensile strength specification.
FlexLab (FLX); Version: FLX-217-40;
Inpeco S.A.
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
IntelePACS - InteleConnect / TechPortal
INTELERAD MEDICAL SYSTEMS INCORPORATED
Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.
FlexLab (FLX); Version: FLX-217-10;
Inpeco S.A.
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
Pivotal Health Solutions
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
Products may contain contamination, which may result in a darker or brown media color.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
MEDLINE convenience kits labeled as: CIRCUMCISION PACK, REF DYNJ16826O
Medline Industries
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
MEDLINE convenience kits labeled as: 1) MINOR EYE TRAY #81, REF DYNJ27466S; 2) MINOR EYE TRAY #81, REF DYNJ27466T.
Medline Industries
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
MEDLINE convenience kits labeled as: 1) CENTRAL LINE PACK, REF DYNJ00281P; 2) FREE FLAP PACK, REF DYNJ68213C.
Medline Industries
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
MEDLINE convenience kits labeled as: GP-LEEP PACK-LF, REF PHS396969004F
Medline Industries
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.