A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.
Other Medical Devices
🏥 Medical Devices • 15,203 recalls
CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US
Graphic Controls Acquisition
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
Atellica CH Calcium_2 (CA_2), Siemens Material Number 11097644
Siemens Healthcare Diagnostics
Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.
Due to an increase in complaint trend for leaks and intermittent/loose connections.
ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1
Varian Medical Systems Imaging Laboratory GmbH
Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location
Due to an increase in complaint trend for leaks and intermittent/loose connections.
If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.
MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Shimadzu Medical Systems
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Due to failure of microbial testing for total aerobic microbial count (TAMC).
A kit component, the PDI Prevantics Swab was recalled due to uncertain validation of test methods used to manufacture the products.
Alinity m HBV AMP Kit (US and CE)
Abbott Molecular
There is a potential for misquantitation high results for negative samples.
Revolution Apex, Revolution CT with Apex Edition
GE Healthcare
The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
Brius Patient Specific Brackets (components in a set of custom metal orthodontic devices); BRIUS , RX Only, Non-Sterile, Custom Made Device, ID: 52345400UL
Brius Technologies
Pontics, brackets, and customized orthodontic appliances were sold and distributed prior to completing the standard design control process, being documented, and released.
MODEL: X-RAY TV SYSTEM SONIALVISION G4
Shimadzu Medical Systems
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
A kit component, the PDI Prevantics Swab was recalled due to uncertain validation of test methods used to manufacture the products.
Philips, REF M5071A, Adult. Smart PADS CARTRIDGE, 8+ years FOR USE WITH HEARSTART HS1 DEFIBRILLATORS
Philips North America
There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.
Surgicare Neuro Pack - Neurological Kit
Stradis Medical, LLC
The outer bag may be perforated or torn, which may result in a breach in the sterility of the kit.
Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach Catalog Number/GPN: IGTCFS-65-1-FEM-TULIP G52917; IGTCFS-65-1-JUG-TULIP G52916; IGTCFS-65-1-UNI-TULIP G52918
Cook Medical Incorporated
IFU update includes an update to the Device description, Intended Use/Indications for Use, Contraindications, Warnings, Precautions, Potential adverse events
Philips, REF M5072A, INFANT/CHILD - SMART PADS CARTRIDGE (0-8 YEARS). Smart PADS CARTRIDGE , (0-8 YEARS) <55 lbs/25 kg, CE 0123, NON-STERILE, Rx only FOR USE WITH HEARSTART DEFIBRILLATORS
Philips North America
There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.
Cook Celect PlatinumVena Cava Filter Set Catalog Number/GPN: IGTCFS-65-1-FEM-CELECT-PT G34502; IGTCFS-65-1-JUG-CELECT-PT G34309; IGTCFS-65-1-UNI-CELECT-PT G34505;
Cook Medical Incorporated
IFU update includes an update to the Device description, Intended Use/Indications for Use, Contraindications, Warnings, Precautions, Potential adverse events