Artis zee biplane, Model Number 10094141

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →
Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee biplane MN, Model Number 10094143

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan

Mar 7, 2022 Other Medical Devices Nationwide View Details →

Artis Q floor, Model Number 10848280

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee ceiling, Model Number 10094137

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q.zen biplane, Model Number 10848355

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q biplane, Model Number 10848282

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q.zen ceiling, Model Number 10848354

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee floor, Model Number 10094135

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Merlin 2 PCS MER3700 programmer Model MER3400 software

St. Jude Medical, Cardiac Rhythm Management Division

Class I - Dangerous

Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.

Mar 10, 2022 Other Medical Devices Nationwide View Details →

Artis zeego, Model Number 10280959

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Merlin PCS 3650 programmer Model 3330 software

St. Jude Medical, Cardiac Rhythm Management Division

Class I - Dangerous

Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.

Mar 10, 2022 Other Medical Devices Nationwide View Details →

As a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that resulted in a difference between the set number of rotations and the actual number of rotations in a specific mode when the product was used with our company's M5 handpiece.

Dec 15, 2021 Other Medical Devices View Details →