🏨

Other Medical Devices

🏥 Medical Devices • 15,809 recalls

Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

May 10, 2024 Other Medical Devices Nationwide View Details →

OMTech Desktop Laser Engraver

Rygel Advanced Machines d/b/a Omtech Laser

Class I - Dangerous

Lack of redundant interlocks, missing required labels and certification labels as required by 21 1010.2.

Jun 7, 2024 Other Medical Devices Nationwide View Details →

The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

May 10, 2024 Other Medical Devices Nationwide View Details →

There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.

Apr 15, 2024 Other Medical Devices Nationwide View Details →

According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.

Mar 27, 2024 Other Medical Devices Nationwide View Details →