If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all
Patient Monitors
🏥 Medical Devices • 389 recalls
Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).
Philips Respironics V60/V60 Plus Ventilator, PCBA Part Number: 453561544461 V60 Device Part Numbers 1053617 (US) and 1137276 (Philippines)
Philips Respironics
Power Management PCBAs may malfunction, causing a power failure leading to ventilator loss of function.
Baxter Epiphany Cardio Server E3 ECG Management System Servers with software versions: a) v6.1.x, b) v6.2.x, c) v7.0.x.
Baxter Healthcare
Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of the caliper tool becoming non-responsive, preventing the caliper function from being used temporarily on software versions v6.1.x, v6.2.x, and v7.0.x of the E3 user interface.
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton-MR1, REF: 161010, Intensive Care Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Mallinckrodt One-Way Valve, 22F x 22M, a) Part #91346 (1 piece) and b) Part #50412 (pack of 5) Used with ventilators.
Mallinckrodt Manufacturing
The One-Way Valve, 22F x 22M may stick and prevent or reduce the flow of ventilated air or oxygen.
Carina Sub-Acute Care Ventilator
Draeger Medical
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
HAMILTON C2 Ventilator Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates
Hamilton Medical
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Oxylog 3000 Plus emergency and transport ventilator
Draeger Medical
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
Devices were imported with Japanese Language and NIST Oxygen Gas Fittings and retrofitted, may pose a serious health risk as the gas pathway of these devices may be contaminated.
Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.
Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.