Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Patient Monitors
🏥 Medical Devices • 381 recalls
Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Mallinckrodt One-Way Valve, 22F x 22M, a) Part #91346 (1 piece) and b) Part #50412 (pack of 5) Used with ventilators.
Mallinckrodt Manufacturing
The One-Way Valve, 22F x 22M may stick and prevent or reduce the flow of ventilated air or oxygen.
Carina Sub-Acute Care Ventilator
Draeger Medical
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
HAMILTON C2 Ventilator Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates
Hamilton Medical
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Oxylog 3000 Plus emergency and transport ventilator
Draeger Medical
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
Devices were imported with Japanese Language and NIST Oxygen Gas Fittings and retrofitted, may pose a serious health risk as the gas pathway of these devices may be contaminated.
Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.
Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.
Trilogy Evo Universal Ventilator, Model Number DS2000X11B
Philips Respironics
The Trilogy Evo Universal ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo Universal is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and pulse rate data when integrated with the appropriate accessories. The ventilator is suitable for use in institutional and hospital settings and non-emergency transport settings; for example, wheelchair. It may be used for both invasive and non-invasive ventilation.
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.
Trilogy Evo Universal Portable Electric Ventilator, Model Number DS2000X11B
Philips Respironics
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.
Life2000 Ventilator PKG, Hospital, a) REF BT200011, b) REF BT-20-0011, ventilatory support for the care of individuals
Baxter Healthcare
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Life2000 Ventilator PKG, Hospital, a) REF BT20007, b) REF BT-20-0007, ventilatory support for the care of individuals
Baxter Healthcare
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Life2000 Ventilator System, a) REF BT-20-0002, b) REF BT-20-0002A, ventilatory support for the care of individuals
Baxter Healthcare
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Life2000 Ventilator , REF MS-01-0118, ventilatory support for the care of individuals
Baxter Healthcare
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.