Product may be mislabeled.
Infusion Pumps
🏥 Medical Devices • 4,262 recalls
Product may be mislabeled.
AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US
Fresenius Medical Care Holdings
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock
REF: 08P4377, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL
Abbott Laboratories
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
REF: 4P52-20 Hemoglobin A1c for use with Architect, IVD, CE, R1 1 x 52 mL, R2 1 x 20 mL, A1cDL 2 x 35 mL
Abbott Laboratories
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
REF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL
Abbott Laboratories
There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
AquaC UNO H [120V, 60Hz, 1000VA, Fresenius US-Version: 24-10VA-1 Vivonic Part Number: G02000970-US
Fresenius Medical Care Holdings
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock
Kits contain leaking bottles of isopropyl alcohol
Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD
Randox Laboratories
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
Nexiva with Single BD Connecta Stopcock Version (Pink) Catalog No. 383687 (OUS)
Becton Dickinson & Company
May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.
Soluble Transferrin Receptors Calibrator Series (STFR CAL), Catalog Number TF10161. IVD
Randox Laboratories
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.
The Instructions for Use (IFU, Part Number 750095) provided with the affected lots by the firm contains an error associated with product s CE marking. The CE mark contained in the IFU is followed by the Notified Body designation, British Standards Institution number 2797 (BSI-2797) for product inserts 750095.28 and 750095.29. The erroneous label CE2797 were in the IFU insert revision 750095.28 for the 408 affected lots that were shipped. No product was packed with erroneous insert revision 750095.29. Since the product is currently CE marked as a Class I - self-certified product under the EU In Vitro Diagnostics Directive, the Notified Body designation should not be specified. As part of the corrective action, the firm has notified the customers who received the products with erroneous labeling, and is conducting the replacements with the correct product insert containing the Insert Revision 750095.30
Nexiva with Single BD Connecta Stopcock Version (20 Ga Blue) Catalog No.383682 (OUS)
Becton Dickinson & Company
May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.
Universal Robot UR5 arm, PV010204 subcomponent of Aesculap Aeos Digital Microscope System, PV010
Aesculap Implant Systems
Malfunction of the robotic arm preventing further movement of the robotic arm and requiring restart of the robotic digital microscope system.
YelloPort Elite Universal Seal. For use in laparoscopic procedures.
Surgical Innovations
There is the potential that YelloPort Elite Universal Seals packaging may contain small holes which may affect the sterility of the product.
Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0
Randox Laboratories
Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.
MiniMed 640G Insulin Pump (O.U.S. Version)
Medtronic MiniMed
Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
MiniMed 740G Insulin Pump (O.U.S. Version)
Medtronic MiniMed
Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
Ivenix Infusion System (IIS)
Ivenix
Downstream occlusion alarm is sometimes immediately followed by a pump problem alarm.